Setup Totalizer Display; Setup Totalizer Mode Display; Setup Totalizer Start Flow Display - Aalborg Command Module Operating Manual

Table of Contents

Advertisement

In order to deactivate the audible signal during operation, press any button on
the key pad (NOTE: Alarm Conditions have to be removed or Alarm for the
corresponding channel has to be disabled).
If Valve action is chosen the power to the mass flow controller valve will be shut off,
allowing the valve to mechanically close.
If the Valve was shut off as the result of the Alarm action, in order to go back to
normal operation the Valve mode has to be changed to AUTO (See section e.3.5).
NOTE: Alarm Conditions have to be removed or Alarm has to be disabled prior
any attempts to change the Valve mode.
Press the ENTER button under EXIT on the display, to go back to the Main Display.
e.3.9 Setup Totalizer Display
Before enabling the Totalizer ensure that all totalizer settings are configured properly.
Totalizer Start values have to be entered in %F.S. engineering units. Totalizer will not
totalize until the flow rate becomes equal or more than the Totalizer Start value.
Totalizer Stop values have to be entered in volume / mass based engineering units.
Totalizer condition becomes valid when the totalized volume is equal or more than the
Totalizer Stop value.
From the Setup Channel Display (see section e.3.7) press the Down button under
TOTAL on the display in order to make Totalizer settings. The following Setup Totalizer
Mode Display will appear.
e.3.9.1 Setup Totalizer Mode Display
To change the Totalizer Mode in ENABLED or DISABLED, press the button (UP, DOWN
respectively) directly below the mode shown on the display. Press the ENTER button
under EXIT on the display, to go back to the Main Display. Press the ESC button under
NEXT on the display to go to the Totalizer Start settings. The following Setup Totalizer
Start Flow Display will appear.
e.3.9.2 Setup Totalizer Start Flow Display
Pressing (+) or (-) will increment or decrement Totalizer Start settings per 0.1% F.S.
respectively. If no settings are required for Totalizer Stop value press the ENTER but-
ton under EXIT on the display, to go back to the Main Display. Press the ESC button
24