Ce Compliance; D.1 Ce Compliance - Aalborg Command Module Operating Manual

Table of Contents

Advertisement

FUSE: 2A on input and output power line. When changing, unplug the device from
power source. Replace only with fuse 5mm 2A/250V FF.
DISPLAY: 24x2 LCD dot matrix with backlight; 24x2 Vacuum Florescent display
optional.
READINGS: 0 to 100% scale is standard; the following units of measure are sup-
ported: %F.S., SLPM, SLPH, SCCM, SCCH, SCFM, SCFH, SCMM, SCMH, LBPM,
LBPH, GRPM, GRPH.
ADC/DAC RESOLUTION: 12 bits (0.025%).
DIGITAL COMMUNICATION STANDARD: RS-232.
Optional: Ethernet TCP/IP (HTML Server or TELNET Console).
RS-232 COMMUNICATIONS PARAMETERS: 8 bits, two stop bits, no parity (8,2,N).
RS-232 DATA TRANSFER RATE: 9600 baud.
CONTACT CLOSURES: Two sets of dry contact relay outputs for each channel (0.5A at
125 Vac, 1A at 24 Vdc).
BUZZER: Piezo Audio Indicator at 3.0 KHz, 85 dB activated by Alarm or Totalizer.
DIMENSIONS: Length: 7.75 inch (19.5cm), width: 6.75 inch (17cm), height: 4.5 inch
(11cm).
WEIGHT: 6.00 lbs (1.8kg).
TRANSDUCER INTERFACE CABLE: Flat cable with 15 pin "D" connector on the ends
is standard for AFC series mass flow controllers. Optional round shielded cable is
available with male/female "D" connector ends for customer's mass flow meters or
controllers. [Cable length may not exceed 9.5 feet (3 meters)].
RELAYS PORT CABLE: Optional shielded cable with female 25 pin "D" connector to
connect to command module relays port.
DATA PORT CABLE: Optional shielded cable with male 25 pin "D" connector to con-
nect to command module data port. [Cable length may not exceed 9.5 feet (3
meters)].

d.1 CE Compliance

Any SDPROC are in compliance with the below stated test standards currently accepted.
EMC Compliance with 89/336/EEC as amended;
CE COMPLIANCE: EN 50011:1998 Group 1, Class A August 1999 for emissions; EN 61000-6-1:2001 for immunity.
5