Medivators CER-1 User Manual

Endoscope reprocessing system
Hide thumbs Also See for CER-1:

Advertisement

Quick Links

SYSTEM OPERATION

Run the Reprocessing Cycle

1. Close CER lid and secure using lid locks
2. Cycle Selection: Automatic FULL or LCG/RINSE
3. Confirm Time Settings: Wash and LCG
4. Select START to begin cycle
5. Basin thermometer HLD temperature verification

Complete the Reprocessing Cycle

1. Verify STOP LED now illuminated
2. Alcohol Flushing Port w/air on (Manual, air/10, Start)
3. Record Endoscope Cycle completion in Log
4. Disconnect hook up and remove and store reprocessed endoscope(s) and hook up(s)

End of Day Shutdown

1. Water Supply Valve: End of Day "OFF"
2. Clean basin, lid and gaskets (Use lint-free cloth)

MAINTENANCE

Weekly Maintenance
1. Clean CER reprocessor and basin
(use lint-free cloth)
2. Inspect CER reprocessor and hook-up components
for wear and tear
3. Lubricate O-rings with silicone oil
Monthly (as specified)
1. 1.0 micron water pre-filter (replace every 3 months
or if below 40 PSI)
2. 0.2 micron absolute filter (replace every 6 months
or if below 40 PSI) and perform waterline
disinfection
3. Active Vapor Management System (if applicable):
Charcoal filter (replace every 6 months)
Medivators Website "Resource Center"
Go to: www.medivators.com, Select "Resource Center" and "User Library" for detailed user guides and hook up
matrices, report forms and logs, and product bulletins
Medivators Customer Service Support Toll Free: Phone: 1-800-444-4729 · FAX: 1-866-421-7696
MEDIVATORS
®
, RAPICIDE
®
and INTERCEPT
®
are registered trademarks of Medivators Inc.
All other company or product names referenced are trademarks of their respective owners.
© 2016 Medivators Inc. All rights reserved. 50097-076/B

TROUBLESHOOTING

Troubleshooting
1. Review troubleshooting section in CER User
Manual
2. Service Contract Coverage and Loaner Repair
Program
Endoscope Reprocessing System
FOR FACILITY USE ONLY
Staff Name:
Training Date:
Trainer:
IN-SERVICE GUIDE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CER-1 and is the answer not in the manual?

Questions and answers

Summary of Contents for Medivators CER-1

  • Page 1: Troubleshooting

    Staff Name: Training Date: Medivators Website “Resource Center” Go to: www.medivators.com, Select “Resource Center” and “User Library” for detailed user guides and hook up matrices, report forms and logs, and product bulletins Trainer: Medivators Customer Service Support Toll Free: Phone: 1-800-444-4729 · FAX: 1-866-421-7696 MEDIVATORS ®...
  • Page 2 High-Level Disinfectant (HLD) 3. Locate proper hook up and connect hook up to endoscope ports and basin connection 1. Amount of HLD disinfectant required: 4 gallons for CER-1, 5 gallons for CER-2 2. Loading disinfectant reservoir using Transfer Pump Endoscope Reprocessing Cycle Summary 3.

This manual is also suitable for:

Cer-2

Table of Contents