2. The customer assumes complete responsibility for meeting each of the specifications and requirements set forth in the Site Requirements, to ensure proper installation and performance of the ADVANTAGE PLUS Endoscope Reprocessor(s). Failure of the customer to comply with these Site Requirements and any deviation from the...
The ADVANTAGE PLUS Reprocessor requires a thermostatic mixing valve with a temperature range of 35°C ± 2°C (95°F ± 4°F). It is the responsibility of the facility to purchase, install and maintain this device. PRODUCT BULLETIN:...
Page 5
• A vented drain system capable of draining 3.17 GPM, 18” or less from the floor will ensure the proper operation of the ADVANTAGE PLUS Endoscope Reprocessor. The supplied drain line has a 1” ID, 1 3/8” OD and is 3 feet long. Any additional required length will be customer supplied.
• A 1/4” ID flexible air hose (EPDM, braided, 150 psi minimum) is required to connect the ADVANTAGE PLUS Endoscope Reprocessor to the house air. The hose connects to a barbed fitting on the back of the ADVANTAGE PLUS Endoscope Reprocessor This is not supplied with the ADVANTAGE PLUS Endoscope Reprocessor.
• If the optional air compressor is used it may be placed on the floor on either side or the back of the ADVANTAGE PLUS Endoscope Reprocessor. This will add additional depth (~16”) or width (~16”) to the installation. The physical dimensions of the compressor are 18”...
Page 8
SITE REQUIREMENTS Figure 1. Front & Back Dimensions Figure 2. Side Dimensions & Clearance...
Page 9
SITE REQUIREMENTS APPENDIX A USER / OPERATOR ID FOR MEDIVATORS REPROCESSING SUITE of EQUIPMENT Group Barcode # A=Administrator PRINT First Facility from badge O=Operator and Last Name (if applicable) B=Biomed (if applicable) D-Doctor (if using)
Page 10
SITE REQUIREMENTS APPENDIX B FACILITY ENDOSCOPE LIST Customer Model # Serial # Brand Internal ID Barcode Department (if applicable)
Page 11
APPENDIX C SITE REQUIREMENTS CHECKLIST This checklist should be utilized to ensure the installation site meets the required specifications. If all specifications are met, the site is suitable for the installation and use of the ADVANTAGE PLUS Endoscope Reprocessor. ®...
Page 12
SITE REQUIREMENTS APPENDIX D JUN-AIR Compact Oil Free Compressor...
Page 13
Endoscope Reprocessors and should not ® be used. The ADVANTAGE PLUS Endoscope Reprocessors require a water supply that is tempered to 35°C ± 2° (95°F ± 4°F) with water at this temperature being available on demand for proper reprocessor operation. Reprocessor...
Page 14
SITE REQUIREMENTS APPENDIX F Endoscope Reprocessor IT FAQ Medical Equipment Configuration Question Answer What CPU is used in the medical device? Intel i3-2120 CPU@3.3 GHz How many processors are included in the medical device? What OS does the system utilize? Windows 7 Can critical security patches be installed without prior vendor approval? Does the device incorporate a network switch into its design?
Question Answer Is local area network connection required for normal operation? YES, on-board Ethernet 1 Gbps, RJ45 Is PC network hardware included in the ADVANTAGE PLUS ® socket. Can unutilized ports be closed without negatively impacting device operation? YES, fixed and dynamic addressing is Can the device support DHCP for network address configuration? supported.
Page 16
Pressure regulators for hot and cold supply Shut off valves for supply lines Drain must be lower than 18 inches from the floor. ADVANTAGE PLUS Reprocessor is gravity drained and will not drain if it is higher than 18 inches. Example of combined temp/pressure...
Page 17
Water consumption is approximately 8 gallons per standard cycle and 13 gallons per cycle if using the detergent wash function • Water hardness must be less than 12gpg or 200ppm • De-ionized water can NOT be used in the ADVANTAGE PLUS Endoscope ® Reprocessor •...
Page 18
SITE REQUIREMENTS APPENDIX H Valve Tag Identification Chart COLD WATER INLET N.O. HOT WATER INLET N.O. SYSTEM DRAIN N.C. MIXING VALVE OUTLET N.O. TEMPERATURE GAUGE...
Page 20
SITE REQUIREMENTS Cantel Medical Devices (China) Co. Ltd. Medivators Inc. Medivators BV Cantel Medical Asia/Paci c Pte. Ltd. Unit 804-805, Innov Tower Block A, 14605 28th Avenue North Sourethweg 11 1A International Business Park Hongmei Road, Xuhui 200233 Shanghai Minneapolis, MN 55447 USA...
Need help?
Do you have a question about the ADVANTAGE PLUS and is the answer not in the manual?
Questions and answers