Omega BB701 Quick Start Quide

Omega BB701 Quick Start Quide

Blackbody calibrator 230 vac model only
Hide thumbs Also See for BB701:

Advertisement

Quick Links

WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects
in materials and workmanship for a period of 13 months from
date of purchase. OMEGA's WARRANTY adds an additional one (1)
month grace period to the normal one (1) year product warranty
to cover handling and shipping time. This ensures that OMEGA's
customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for
evaluation. OMEGA's Customer Service Department will issue an
Authorized Return (AR) number immediately upon phone or written
request. Upon examination by OMEGA, if the unit is found to be
defective, it will be repaired or replaced at no charge. OMEGA's
WARRANTY does not apply to defects resulting from any action of
the purchaser, including but not limited to mishandling, improper
interfacing, operation outside of design limits, improper repair,
or unauthorized modification. This WARRANTY is VOID if the unit
shows evidence of having been tampered with or shows evidence of
having been damaged as a result of excessive corrosion; or current,
heat, moisture or vibration; improper specification; misapplication;
misuse or other operating conditions outside of OMEGA's control.
Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various
products. However, OMEGA neither assumes responsibility for
any omissions or errors nor assumes liability for any damages
that result from the use if its products in accordance with
information provided by OMEGA, either verbal or written.
OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA
MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF
ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT
THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING
ANY WARRANTY OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth
herein are exclusive, and the total liability of OMEGA with
respect to this order, whether based on contract, warranty,
negligence, indemnification, strict liability or otherwise,
shall not exceed the purchase price of the component upon
which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used,
nor shall it be used: (1) as a "Basic Component" under 10 CFR 21 (NRC),
used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or
with any nuclear installation or activity, medical application, used on
humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/DISCLAIMER language, and,
additionally, purchaser will indemnify OMEGA and hold OMEGA
harmless from any liability or damage whatsoever arising out of the use
of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA
Customer Service Department.
BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM OMEGA'S CUSTOMER
SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS). The assigned AR number should then be marked on the
outside of the return package and on any correspondence.
FOR WARRANTY RETURNS,
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair
please have the following
charges. Have the following
information available BEFORE
information available BEFORE
contacting OMEGA:
contacting OMEGA:
1. Purchase Order number under
1. Purchase Order number to
which the product was
cover
PURCHASED,
the COST of the repair or
2. Model and serial number of
calibration,
the product under warranty, and
2. Model and serial number of the
3. Repair instructions and/or
product, and
specific problems relative to
3. Repair instructions and/or specific
problems relative to the product.
the product.
OMEGA's policy is to make running changes, not model changes,
whenever an improvement is possible. This affords our customers
the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2013 OMEGA ENGINEERING, INC. All rights reserved.
This document may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine-readable
form, in whole or in part, without the prior written consent of OMEGA
ENGINEERING, INC.
MQS3264/0813
5
1. Press the
key to enter the programming mode.
The lower display will alternately display the menu
level and "Ac.Cd."
2. Use the
and
keys to change to the desired
menu level.
3. Once you have chosen the desired menu use the
key to scroll through the parameters. To change the
setting of a given parameter, use the
and
keys.
4. To save settings press the
key. The controller
now exits the programming menu and return to the
normal operating mode.
5. To change settings on other menu levels, you must
re-enter the programming menu (from step #1).
Target Plate Air Purge
The BB701 comes with a nitrogen air purge collar
that surrounds the target plate. When connected to
a nitrogen gas source the target plate is kept free of
condensation at lower temperatures. Snow is likely to
form on the plate when it is kept at temperatures below
freezing temperature. Snow or condensation will change
a target's emissivity. The can lead to inaccuracies unless
the infrared pyrometer is corrected for this emissivity
change. It is therefore recommended that the user take
advantage of this feature.
Use dry gas or Nitrogen with a dew point below -20°C
(-4.0°F). The rate of gas inlet should be 0.25 to 0.50 m
hr. (5 to 10 ft
3
/hr.). Do not exceed the maximum gas
inlet rate. The port will accept any flexible tubing (vinyl,
rubber, Teflon, latex) with a ¼" outer diameter, such as
OMEGA
®
#TYVY-1418-100-A.
Reference Probe
An RTD probe has been embedded into the target plate
heater assembly to be used as a reference to check the
calibrator's calibration and accuracy. A connector has
been provided with your calibrator for connection to
this reference probe. Listed below are a few calibration
points and the resistance readings you can expect from
the reference probe.
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
6
Target Plate Temperature Reference Probe Resistance
-10°C (14°F)
96.09 Ohms
+10°C (+50°F)
103.90 Ohms
+50°C (+122°F)
119.40 Ohms
+70°C (+158°F)
127.07 Ohms
+100°C (+212°F)
138.05 Ohms
+120°C (+248°F)
146.06 Ohms
TARGET PLATE ASSEMBLY
BLACK WIRE
RED WIRE
BLACK WIRE
RTD REFERENCE PROBE
Internal Reference Probe Connections
RS232 Communication
The RS232 communications port allows bi-directional
data transfer via a three conductor cable consisting
of signal ground, receive input, and transmit output.
It is recommended that less than fifty feet of cable be
used between the computer and this instrument. Use a
2
/
straight DB9 (female) to DB9 (male) connector cable to
connect your computer to the BB701. The cable should
be attached only when the computer and BB701 are off.
1
2
3
BB701
BLACKBODY
CALIBRATOR
6
7
8
COMPUTER
SERIAL
1
2
3
PORT
(RS232)
6
7
8
Internal Reference Probe Connections
For complete product manual:
www.omega.com/manuals/manualpdf/M3264.pdf
REAR PANEL
BB701
1
3
2
Blackbody Calibrator
REFERENCE
PROBE
CONNECTOR
4
5
9
omega.com info@omega.com
Servicing North America:
U.S.A.:
OMEGA Engineering, Inc., One Omega Dr.
P.O. Box 4047, Stamford, CT 06905-0047 USA
Toll-Free: 1-800-826-6342 (USA & Canada Only)
Customer Service: 1-800-622-2378 (USA & Canada Only)
Engineering Service: 1-800-872-9436 (USA & Canada Only)
Tel: (203) 359-1660
e-mail: info@omega.com
For Other Locations Visit omega.com/worldwide
4
5
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety
and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of
9
its products to the European New Approach Directives. OMEGA will add the CE
mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but
OMEGA accepts no liability for any errors it contains, and reserves the right to
alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used
for, human applications.
230 VAC Model Only
®