Frequency compensation
Before taking any measurements using the probe, first check its
compensation and adjust it to match the channel inputs.
Connect the probe to a 2 V pk-pk, 1 kHz square wave source. Most PicoScope
oscilloscopes have a signal generator output marked GEN or AWG, which you
can configure to generate such a signal. Set the switch on the probe to the
X10 position. Adjust the trimmer until you see a flat-top square wave on the
display:
INCORRECT
Accessories and features
The probe is provided with several accessories designed to make probing
and measurement simpler. Please take a moment to familiarize yourself with
these accessories and their uses.
Accessories included
Retractable hook tip
Ground lead
Adjustment tool
Tip insulating sleeve
Cable marker

Optional accessories

Order code
Description
TA384
Replacement rigid probe tips for TA375 and TA386, pack of 5
TA385
Replacement spring probe tips for TA375 and TA386, pack of 5
INCORRECT
TA375
1
1
1
1
2
CORRECT
TA386
1
1
1
1
8