Download Print this page

PICO TA386 User Manual page 4

100/200 mhz oscilloscope probe

Advertisement

Voltage v frequency rating curve
700
600
500
400
300
200
100
0
10 kHz
WARNING
This probe does not carry a measurement category rating. To prevent
electric shock, do not connect to a mains (line) voltage or a derived
voltage that can carry the overvoltage transients that may be present.
Measurement categories are defined in IEC 61010-031 as follows:
No Measurement Category (not in CAT II, III or IV)
Definition
For measurements performed on circuits not directly connected
to a mains supply.
Grounding
WARNING
Never connect the ground input to, or allow it to touch, any electrical
potential other than ground. To prevent personal injury or death, use a
voltmeter to check that there is no significant AC or DC voltage between
the probe ground and the point to which you intend to connect it.
CAUTION
Applying a voltage to the ground input is likely to cause permanent
damage to the probe or other connected equipment.
It is good practice to connect the probe output to the measurement instrument
and the ground lead to earth ground before connecting the probe to the circuit
under test. Disconnect the probe input and the probe ground lead from the
circuit under test before disconnecting the probe from the measurement
instrument.
100 kHz
1 MHz
10 MHz
100 MHz
1 GHz

Advertisement

loading

This manual is also suitable for:

Ta375