How To Create a Program
The ultrasonic processor has the ability to save up to 10 programs.
1.
Select a program number from the keypad, the program number w ill appear above the
Amplitude box.
For each of the following steps: After touching a field, the background w ill change to yellow
indicating an active field. After entering data, the ent ry is completed by touching Enter (ENT),
touching the field a second time or by touching a new field.
2.
Program a value into the Amplitude field.
3.
Program total Process Time (total active sonication time).
Program Pulse On time – If no data is entered (00hr:00min:00sec), the unit w ill run
4.
continuously w ithout pulsing.
Program Pulse Off time – If no data is entered, the unit w ill run continuously w ithout pulsing.
5.
Process Time, Pulse On and Pulse Off times are measured in Hours: Minutes: Seconds format.
6.
Activate Temperature Monitoring (if necessary).
7.
Program a Shutdow n Temperature (if necessary). Enter a temperature value.
During sonication, if the temperature reaches the Shutdow n value, the unit w ill pause
sonication. The unit w ill remain paused until the temperature drops below the pre-set value.
Once the temperature drops below the value, sonication can be manually resumed.
8.
Touch Save to store the program to the memory number selected in step 1.
9.
Touch Run to proceed to the follow ing screen.
Program
Progress
10. Touch Start to begin the program.
Programming Screen
20
Total programmed time of
active sonication. Rest or
Off time is not included in
this value.
Total time of active
sonication that has occurred
during the current program.
Rest or Off time is not
included in this value.