Qsonica Q700 Operation Manual page 16

Sonicator ultrasonic processor
Hide thumbs Also See for Q700:
Table of Contents

Advertisement

fluctuation in w attage readings. Always use the wrenches supplied with the unit.
6. If you w ill be using a Microtip or extender, remove the flat tip on the end of the
replaceable tip probe, t hen attached the Microtip or extender in its place.
7. Horns and probe tips w ear after normal usage. Using a severely w orn probe tip can
damage internal generat or components.
8. If using a laboratory stand, mount the convertor /probe assembly using a clamp. Be
sure to secure the clamp to the upper section of the convertor housing only. Never
secure the clamp to any other portion of t he convertor/probe assembly. If you are using
an acoustic enclosure mount the convertor properly in the convertor collar.
9. Connect t he converter cable to the pow er supply and then to the top of the convertor.
Push the connectors in and turn the chrome rings clockwise ¼ turn to secure the
connectors.
10. If application or procedure instructs that t he unit be run for a period of time that w ill
allow the probe/convert or assembly to be w arm to t he touch, it is recommended that
the convertor be air cooled w ith dry compressed air. Never run the convertor and horn
assembly if it becomes hot. Connect either one of the air fittings on top of the
converter to a source of dry compressed air. This will cool the convertor so that it can
function normally under load for an extended duration without becoming too hot.
Operation:
Your new ultrasonic processor has been designed w ith a color LCD user interface w it h
touch screen capabilities. All program and run functions are controlled through the touch
screen panel.
This is the first screen t hat appears aft er Ultrasonic processor is sw itched On.
1. Answ er the Microtip Question, Yes or No.
When using a Microtip, the YES button must be selected.
16
Rev. 6-12

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents