Using Supplemental Oxygen - ResMed Flow Generator AutoSet CS 2 User Manual

Resmed flow generator user guide
Hide thumbs Also See for Flow Generator AutoSet CS 2:
Table of Contents

Advertisement

Available languages

Available languages

U
S
SING
UPPLEMENTAL
If oxygen is being added at the mask, the oxygen tube is attached to the mask port.
Refer to your mask manual for full instructions.
Up to 15 L/min of oxygen can be added at the mask when using an A
system.
Oxygen tube
Figure 11: Oxygen tube attached to Mirage FFMS2 and Ultra Mirage masks
WARNINGS
!
If oxygen is used with this device, the oxygen flow must be turned off when the
device is not operating.
Explanation: When the device is not in operation, and the oxygen flow is left
on, oxygen delivered into the ventilator tubing may accumulate within the
device enclosure and create a risk of fire.
Always begin A
Always turn the oxygen supply off before stopping A
Oxygen supports combustion. Oxygen should not be used while smoking or in
the presence of an open flame.
O
XYGEN
CS2 therapy before the oxygen supply is turned on.