IEC 60601-1 Classifications
Class II (double insulation)
Type CF
Table 7: Displayed values
Value
Pressure sensor at air outlet
Pressure
Flow sensor in flow generator*
Leak
Tidal volume
Respiratory rate
Minute ventilation
*
** The displayed values are estimates. They are provided for trending purposes only.
Pressure Variation
25.00
20.00
15.00
10.00
5.00
0.00
36
Range
-5 to 30 cm H
0–120 L/min
100–3000 mL
6–60 BPM
0.6–60 L/min
Results may be inaccurate in the presence of leaks.
0
Accuracy
O
±0.5 cm H
2
**
**
±0.5 BPM
**
1
Breaths
Display Resolution
O
0.1 cm H
O
2
2
1 L/min
1 mL
0.1 BPM
0.25 L/min
Max Pressure
2/3 Max Pressure
1/3 Max Pressure
10 BPM
15 BPM
20 BPM
2
3

Hide quick links: