Table of Contents

Advertisement

Subsystem Syntax
:ERRor?
Comments
Example
:VERSion?
Comments
The SYSTem command subsystem returns error numbers and their associated
messages from the error queue. You can also query the SCPI version to which this
instrument complies.

SYSTem

:ERRor?
:VERSion?
SYSTem:ERRor? returns the error numbers and corresponding error messages in
the error queue. See Appendix B in this manual for a listing of the error numbers,
messages and descriptions.
When an error is generated by the digitizer, it stores an error number and
corresponding message in the error queue.
One error is removed from the error queue each time the SYSTem:ERRor?
command is executed. The errors are cleared in a first-in, first-out order. This
means that if several errors are waiting in the queue, each SYSTem:ERRor?
query returns the oldest (not the most recent) error. That error is then removed
from the queue.
When the error queue is empty, subsequent SYSTem:ERRor? queries return
+0,"No error". To clear all errors from the queue, execute the *CLS
command.
The error queue has a maximum capacity of 20 errors. If the queue overflows,
the last error is replaced with -350,"Too many errors". No additional
errors are accepted by the queue until space becomes available.
Reading the Error Queue
SYST:ERR?
enter statement
SYSTem:VERSion? returns the SCPI version number this instrument complies.
The information returned is in the format "YYYY.R" where "YYYY" is the year and
"R" is the revision number within that year.
SYSTem
Query the error queue
Enter readings into computer.
Digitizer Command Reference
87