*CLS
Comments
*ESE and *ESE?
Parameters
Comments
Example
*CLS clears the Standard Event Status Register, the Operation Status Register, the
Questionable Signal Register, and the error queue. This clears the corresponding
summary bits (3, 5 and 7) in the Status Byte Register. *CLS does not affect the
enable unmasks of any of the Status Registers.
Executable when initiated: No
Coupled command: No
Related Commands: STATus:PRESet
Reset (*RST) condition: None
*ESE < unmask > enables (unmasks) one or more event bits of the Standard Event
Status Register to be reported in bit 5 (the Standard Event Status Summary Bit) of
the Status Byte Register. unmask is the sum of the decimal weights of the bits to be
enabled allowing these bits to pass through to the summary bit ESB (bit 5 in the
status byte).
The query form returns the current enable unmask value.
Parameter
Name
unmask
A 1 in a bit position enables the corresponding event; a 0 disables it.
Executable when initiated: Yes
Coupled command: No
Related Commands: *ESR?, *SRE, *STB?
Reset (*RST) condition: unaffected
Power-On condition: no events are enabled
Enable all error events
*ESE 60
Parameter
Range of
Type
numeric
0 through 255
Enable error events