Probable Causes
Suggested Actions
Troubleshooting and Test Functions
Pump Configuration
Pump Configuration
At switch-on, the pump has recognized a new pump configuration.
The binary pump is assigned its configuration at the factory. If the active-inlet
valve and pump encoder of channel B are disconnected, and the binary pump
is rebooted, the error message is generated. However, the binary pump will
function as an isocratic pump in this configuration.The error message
reappears after each switch-on.
• Active-inlet valve and pump encoder of channel B disconnected.
Reconnect the active-inlet valve and pump encoder of channel B.
69