Fluke VT650 User Manual page 53

Gas flow analyzer
Table of Contents

Advertisement

Parameter
Inspiratory Time
(The time of the inspiratory period
including the inspiratory hold time.)
Inspiratory Hold Time
(The time of the End Inspiratory State.)
Expiratory Time
(The time of the expiratory period
including the expiratory hold time.)
Expiratory Hold Time
(The time of the End Expiratory State.)
Peak Inspiratory Flow
(The maximum of the absolute value of
the Flow Rate measured during the
Inspiratory Time.)
Peak Expiratory Flow
(The maximum of the absolute value of
the Flow Rate
measured during the Expiratory Time.)
[1] Inspiratory pause time >0.5 sec
Table 10. Breath Parameter Accuracy (cont.)
Abbreviation
Resolution
Ti
0.01 sec
TiH
0.01 sec
Te
0.01 sec
TeH
0.01 sec
PIF
0.01 lpm
PEF
0.01 lpm
Range
0 sec to 60 sec
0.02 sec
0 sec to 60 sec
1 % or 0.1 sec
0 sec to 90 sec
0.5 % or 0.01 sec
0 sec to 90 sec
0.02 sec
±300 lpm
±1.7 % or 0.04 lpm
±300 lpm
±1.7 % or 0.04 lpm
Gas Flow Analyzer
Specifications
Accuracy
45