VT650/VT900
Users Manual
Gas Correction
ATP
Ambient temperature and pressure (actual humidity)
ATPD
Ambient temperature and pressure, dry (0 % humidity)
ATPS
Ambient temperature and pressure, saturated (100 % humidity)
STP20
Standard temperature 20 °C, standard pressure 760 mmHg (actual humidity)
STP21
Standard temperature 21 °C, standard pressure 760 mmHg (actual humidity)
STPD0
Standard temperature 0 °C, standard pressure 760 mmHg, dry (0 % humidity)
STPD20
Standard temperature 20 °C, standard pressure 760 mmHg, dry (0 % humidity)
STPD21
Standard temperature 21 °C, standard pressure 760 mmHg, dry (0 % humidity)
BTPS
Body temperature 37 °C, ambient pressure 760 mmHg, saturated (100 % humidity)
BTPD
Body temperature 37 °C, ambient pressure 760 mmHg, dry (0 % humidity)
Note: Gas correction applies to the airflow channel. The ultra-low-flow channel of the VT900 is set to STPD21.
30
Table 6. Gas Correction Mode