Instrument Compatibility; Measurement Considerations - Fluke 80i-500s Instruction Sheet

Ac current probe for oscilloscopes
Table of Contents

Advertisement

Maximum Conductor Size:
One 30 mm (1.18 in.), 750 kcmils (MCM) THHN, two 25 mm
(0.98 in.), 500 kcmils (MCM) THHN, or one 63 x 5 mm (2.5 x 0.2
in.) Bus Bar.
Output Cable: 1.6 meters (63 in.), BNC termination. Cable and
BNC are rated to 600V ac rms for enhanced safety.
Mechanical Shock: 100G per IEC 68-2-27

INSTRUMENT COMPATIBILITY

The 80i-500s is compatible with any oscilloscope, voltmeter, or
other voltage measuring device that has the following features:
Accepts a standard BNC connector. A BNC-to-bannana
adapter (order PM 9053 from Fluke) can also be used with
standard inputs on a digital multimeter (DMM).
Range and resolution capable of displaying 1mV of output
per amp of measured current.
AC Volts accuracy of 1% or better to take full advantage of
the accuracy of the probe.
Input impedance of greater than or equal to 1MΩ in parallel
with a maximum of 47 pF.
OPERATING INSTRUCTIONS (Scope Mode)
1.
Connect the 80i-500s Current Probe to the desired input
channel on the oscilloscope.
2.
Use a 1 : 1 probe setting.
3.
Clamp the current probe around the conductor.
Press S. to select Scope Mode.
4.
Press A.
5.
Adjust V or v (volts/division) for the best vertical display.
6.
(For line power currents less than 100A, 50 mV/div is a good
first choice.)
Adjust t (time/division) for the best horizontal display.
7.
(For power line frequency, 5 msec/div is a good choice.)
OPERATING INSTRUCTIONS (Meter Mode)
1.
See Figure 3. Connect the 80I-500s Current probe to
Channel A.
2.
Use a 1 : 1 probe setting.
3.
Clamp the current probe around the conductor.
Press Mto select the Meter Mode.
4.
Press u or d to select an appropriate meter range.
5.
(The current probe output is 1 mV per ampere measured.)

MEASUREMENT CONSIDERATIONS

Observe the following guidelines for positioning the current probe
jaws:
Center the conductor inside the probe jaws.
Make sure the probe is perpendicular to the conductor.
If possible, avoid measurements close to other current
carrying conductors.

Advertisement

Table of Contents
loading

Table of Contents