Specification
Details
Over voltage
2 (according to standard EN 61010-1)
category
Pollution degree
2 (according to standard CEI 664)
Altitude
< 2000 m
Measurement
Cat II, Class 1 (overvoltage < 1500V)
category
Maximum panel
850 x 450 x 252.5mm [33.46 x 17.71 x 9.94in]
dimension (H x L x
D)
Inlet
Simple fittings for 6 mm O.D. tubing or ¼" O.D. in PE-low density. ¼" OD in PHED-PTFE-
SS as option
Outlet
Barbed stem for 12 mm (½" I.D.) hose
Protection rate
Transmitter: IP65 (NEMA 4)
Panel: IP50 (dust protection)
Optional Enclosure: IP54 (splash water proof)
Instrument is designed to avoid DIPA vapor inside the enclosure. All DIPA vapor is
collected and sent to the instrument drain
Cell
PMMA - compact (minimum tubing)
Flame rate
Conform UL
Maximum weight
15 - 30 Kg
Mains power
100 - 240 VAC, 50-60 Hz, ± 10%, automatic switching
supply
Max. consumption
80 VA
Fuse
5 x 20 cartridge T2AL-250V following CEI127
Display
Curve trend, Last Cal Date, Historical, Concentration, Temperature, Potential
Analog outputs
Number: 6
4-20 or 0-20 mA (650 ohms)
Linear / Dual / Logarithm
Event indication
Relays
4 x Relay (conc)
1 x Warning
1 x System
Logic input
Active / Inactive channels
Remote AutoCal
EMC requirements
EN61326-1: EMC Directive
Note: This is a Class A product. In a domestic environment this product may cause radio interference in
which case the user may be required to take adequate measures.
Measurement variations of less than 5% of the full range can occur if the instrument is
subject to a strong electromagnetic field.