4.1

Operating the T-620/T-620H pump.

Warning
Do not attach the pump to a high pressure source. Follow
these instructions carefully to avoid damage to the pump
and/or personal injuries.
This routine must be followed in order to operate the system correctly
(see fig. 1 accordingly).
1.
If the pump has been ordered alone without system C, the
reservoir will be empty. Remove the fill plug (ref. 1) on the
top of cap and fill the reservoir approximately 2/3 full of fluid.
Caution...
Do not use solvents or synthetic fluids that will adversely
affect the operation of the pump.
Connect and tighten fill plug.
Ordering system C the reservoir will be supplied with the
fluid ordered from our factory.
2.
Remove plug (ref. 2) and attach gauge or pressure device to
be calibrated to the 1/4" BSP hose fitting (ref. 3). Tighten
threads loosely.
Note...
After connection of all parts and before pressure is created,
the air can be released from the system by loosening the
connectors (pos. 3). It is now possible to pump out oil and
air until no air is left in the oil.
3.
Turn needle valve (ref. 4) on left side of the pump counter
clockwise until fully open.
4.
Pump the movable handle (ref. 5) until the test fluid is
evident at the thread connection.