Adjustment Steps - Fluke 8808A Calibration Manual

Digital multimeter
Hide thumbs Also See for 8808A:
Table of Contents

Advertisement

8808A
Calibration Manual
Step
Func./Command
Volts DC: zero points using the 4-wire short
1
VDC
Volts DC: Gains – Adjust using 5700A/5725A
2
VDC
3
VDC
4
VDC
5
VDC
6
VDC
7
VDC
8
VDC
9
VDC
10
VDC
11
VDC
12
VDC
13
VDC
14
VDC
15
VDC
16
VDC
17
VDC
18
VDC
19
VDC
20
VDC
21
VDC
Volts AC: Zero points – Adjust with 5700A in standby or disconnected
22
VAC
Volts AC: Gains – Adjust using 5700A/5725A
23
VAC
24
VAC
25
VAC
26
VAC
27
VAC
28
VAC
29
VAC
30
VAC
3-12
Table 3-9. Adjustment Steps
Point
Input Signal
0 mV/0V
0 mV
-199.9 mV
-199.9 mV
-100 mV
-100 mV
100 mV
100 mV
199.9 mV
199.9 mV
-1.999 V
-1.999 V
-1 V
-1 V
1 V
1 V
1.999 V
1.999 V
-19.99 V
-19.99 V
-10 V
-10 V
10 V
10 V
19.99 V
19.99 V
-199.9 V
-199.9 V
-100 V
-100 V
100 V
100 V
199.9 V
199.9 V
-1000 V
-1000 V
-500 V
-500 V
500 V
500 V
1000 V
1000 V
0 mV/0 V
0 mV
10 mV
10 mV@1 kHz
100 mV
100 mV@1 kHz
150 mV
150 mV@1 kHz
199.9 mV
199.9 mV@1 kHz
0.1 V
0.1 V@1 kHz
1 V
1 V@1 kHz
1.5 V
1.5 V@1 kHz
1.999 V
1.999 V@1 kHz
Description
Zero point of VDC, all ranges
Gain of VDC, 200 mV range
Gain of VDC, 2 V range
Gain of VDC, 20 V range
Gain of VDC, 200 V range
Gain of VDC, 1000 V range
Zero point of VAC, all ranges
Gain of VAC, 200 mV range
Gain of VAC, 2 V range