Introduction
The 85RF II High Frequency Probe (the Probe) transforms a
dc voltmeter into a high frequency (100 kHz to 500 MHz) ac
voltmeter. Conversion from ac to dc is done on a one-to-one
basis and includes a range of 0.25 V rms to 30 V rms. The
Probe's dc output is calibrated to be equal to the rms value
of a sine wave input.
Safety
Symbol
W
X
~
To prevent possible electrical shock, fire, or personal
injury:
Examine the case before you use the Product.
Look for cracks or missing plastic. Carefully
look at the insulation around the terminals.
Limit operation to the specified measurement
category, voltage, or amperage ratings.
Do not touch voltages >30 V ac rms, 42 V ac
peak, or 60 V dc.
Do not use the Product around explosive gas,
vapor, or in damp or wet environments.
Do not use and disable the Product if it is
damaged.
Use the Product only as specified, or the
protections supplied by the Product can be
compromised.
Do not exceed the Measurement Category (CAT)
rating of the lowest rated individual component
of a Product, Probe, or accessory.
Keep fingers behind the finger guards on the
Probes.
Do not connect directly to mains.
PN 4096837
September 2011, Rev. 1, 12/11
©2011 Fluke Corporation, Printed in the U.S.A. All product names are
trademarks of their respective companies. Specifications are subject to change
without notice.
High Frequency Probe
See Instruction Sheet.
Hazardous voltage.
Do not dispose of this product as
unsorted municipal waste. Go to Fluke's
website for recycling information.
WXWarning
85RF II
Instruction Sheet
Description
®

Advertisement

loading

Summary of Contents for Fluke 85RF II

  • Page 1 High Frequency Probe Instruction Sheet Introduction The 85RF II High Frequency Probe (the Probe) transforms a dc voltmeter into a high frequency (100 kHz to 500 MHz) ac voltmeter. Conversion from ac to dc is done on a one-to-one basis and includes a range of 0.25 V rms to 30 V rms. The Probe’s dc output is calibrated to be equal to the rms value...
  • Page 2: Specifications

    Operating altitude: 3000 m Warranty: 1 year Voltmeter Compatibility The 85RF II can be used with a dc voltmeter that has an input resistance of 10 Me, ±10 %. Voltmeters with higher input resistances are usable if an external shunt is put across the input terminals.
  • Page 3: Operation

    Operation To operate the 85RF II Probe: 1. Connect the Probe output to a compatible voltmeter with an applicable shunt (if necessary.) The Probe’s output low must be connected to voltmeter low. 2. Set a voltage range (1 V dc reading for each 1 V rms Probe input).
  • Page 4: Performance Test/Calibration

    If the Probe cannot be calibrated to the specifications in this document, the Probe could be damaged and require replacement or repair. Contact Fluke for instructions. 1. Configure the equipment as shown in Figure 2. Use the leveled sine function of the scope output to set the AC Calibrator signal generator for 1 V rms at 25 MHz and the DMM to measure 1 V dc.
  • Page 5 Fluke 85RF Probe under test Required Equipment List The following equipment (or equivalent) is required for testing the performance of the 85RF II: Equipment Type Recommended Model AC Calibrator Fluke 5520A-SC600 50-Ohm Feed Thru JFW Industries P/N 50L-001 Termination Pomona P/N 1269...
  • Page 6 WXWarning For safe operation and maintenance of the product, remove the input signals before you clean the Product. Use a soft cloth dampened in a weak solution of detergent and water to clean the 85RF II. Do not use solvents.