2-Wire Ohms Verification - Fluke 8808A Calibration Manual

Digital multimeter
Hide thumbs Also See for 8808A:
Table of Contents

Advertisement

2-Wire Ohms Verification

To verify the 2-Wire Ohms function of the Meter, connect it to the test equipment as
shown in Figure 3-3 and apply the resistances listed in Table 3-5.
8808A
INPUT
SENSE
V
2W/4W
HI
1000 V
CAT I
1000V
MAX
750V
600 V
CAT II
1V
LO
500 V pk
200 mA
MAX
10 A
MAX
mA
Nominal
Range
Input
0 Ω
200 Ω
0.008 Ω
190 Ω
200 Ω
STD+0.046 Ω
0 kΩ
2 kΩ
0.00004 kΩ
1.9 kΩ
2 kΩ
STD+0.00033 kΩ
0 kΩ
20 kΩ
0.0004 kΩ
19 kΩ
20 kΩ
STD+0.0033 kΩ
0 kΩ
200 kΩ
0.004 kΩ
190 kΩ
200 kΩ
STD+0.033 kΩ
0 MΩ
2 MΩ
0.00006 MΩ
Connect 57X0A and 5725A Callibrator to the UUT as follows:
4 W
Set 57X0A to EXT Sense and 2-wire Comp On
HI
300V
LO
10A
NC
NC
UUT
Input
HI
Input
LO
Input
HI
Input
LO
Note: Connect the 57X0A sense leads to the UUT first!
Figure 3-3. 2-Wire Ohms Test Setup
Table 3-5. 2-Wire Ohms Verification Steps
90-Day Test Limit
High
-0.008 Ω
[1]
STD-0.046 Ω
-0.00004 kΩ
[1]
STD-0.00033 kΩ
-0.0004 kΩ
[1]
STD-0.0033 kΩ
-0.004 kΩ
[1]
STD-0.033 kΩ
-0.00006 MΩ
Performance Verification and Calibration
57X0A
Output
Output
Sense
Sense
Low
High
0.008 Ω
[1]
STD+0.065 Ω
0.00006 kΩ
[1]
STD+0.0044 kΩ
0.0006 kΩ
[1]
STD+0.0044 kΩ
0.006 kΩ
[1]
STD+0.044 kΩ
0.00008 MΩ

2-Wire Ohms Verification

57X0A
OUTPUT
SENSE
Ω
V A
VA
WIDEBAND
HI
HI
LO
LO
HI
AUX
GROUND
GUARD
CURRENT
HI
LO
HI
LO
ffw002.eps
1-Year Test Limit
Low
-0.008 Ω
[1]
[1]
STD-0.065 Ω
-0.00006 kΩ
[1]
STD-0.0044 kΩ
-0.0006 kΩ
[1]
STD-0.0044 kΩ
-0.006 kΩ
[1]
[1]
STD-0.044 kΩ
-0.00008 MΩ
3
[1]
[1]
3-7