Calibration Adjustment Steps For 1587 Fc/1587 Models With Sn >32980000 - Fluke 1587 FC Calibration Manual

Insulation multimeters
Hide thumbs Also See for 1587 FC:
Table of Contents

Advertisement

1587 FC/1587/1577
Calibration Manual
Table 7. Calibration Adjustment Steps for 1587 FC/1587 Models with SN >32980000
Switch Position
(Function)
AC Volts
DC Volts
DC Millivolts
Ohms
Continuity
Diode Test
milliamps
Insulation
Insulation
Notes:
[1]
Internal HV output calibration. Ensure no lead is inserted in Insulation (+) jack.
34
Input
Terminal
V/COM
mA/COM
mA/COM
To avoid electrical shock hazard, remove all test leads from
the UUT before performing the following step.
[1]
None
Calibration Adjustment Step



























XWWarning

Input Value
600.0 mV, 60 Hz
600.0 mV, 5 kHz
6.000 V, 60 Hz
6.000 V, 5 kHz
60.00 V, 60 Hz
60.0 V, 5 kHz
600.0 V, 60 Hz
600.0 V, 2 kHz
6.000 V, 0 Hz
60.00 V, 0 Hz
600.0 V, 0 Hz
600.0 mV, 0 Hz
60.00 mV, 0 Hz
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
0.000 Ω
50.0 MΩ
600.0 Ω
6.000 V, 0 Hz
60.00 mA, 60 Hz
400.0 mA, 60 Hz
60.00 mA, 0 Hz
400.0 mA, 0 Hz
1.0 mA, 0 Hz
None (press BLUE to
start internal cal)