Error Codes; Table 10 Rfl100 Error Codes - Vaisala RFL100 User Manual

Vainet wireless humidity and temperature data logger
Hide thumbs Also See for RFL100:
Table of Contents

Advertisement

Problem
RFL100 turns off by itself.
6.2 Error Codes
Table 10 RFL100 Error Codes
Error Code
Cause
Err 100
User parameter bank checksum failure.
Err 101
Factory parameter bank checksum
failure.
Err 102
Real-time clock of the data logger has
lost accurate time.
Err 103
Main battery voltage is critically low.
Data logger has stopped radio
communication to conserve energy but
continues to record measurement data in
the local memory.
Err 104
Incorrect factory configuration
parameters.
Err 105
Real-time clock hardware error.
Err 200
Real-time clock battery voltage is low.
Possible Cause
Display and LED of the RFL100
have been turned off remotely
using viewLinc Enterprise
Server. All other functions of
the data logger remain active.
Main batteries are empty.
Chapter 6 – Troubleshooting
Solution
You can change the setting
from the Hosts and Devices
tree in viewLinc Enterprise
Server.
If you are connecting the
RFL100 to a new viewLinc
Monitoring System and you
want the display and LED back
on again, press the Release
button of the RFL100 and
verify that the message NWK
REL appears on screen.
Replace the main batteries. See
Changing RFL100 Batteries
(page
37).
Recommended Action
Turn the data logger off and on again. If
the error persists, contact Vaisala.
Turn the data logger off and on again. If
the error persists, contact Vaisala.
Restore the wireless connection to an
AP10 access point. RFL100 will
synchronize its clock with the time from
the access point.
Replace main batteries. See
RFL100 Batteries (page
Turn the data logger off and on again. If
the error persists, contact Vaisala.
Turn the data logger off and on again. If
the error persists, contact Vaisala.
Replace the clock battery. See
RFL100 Clock Battery (page
Changing
37).
Changing
38).