11. Select menu item [5] Note to edit the calibration info text that is stored in the probe. Edit
the text using the select button and arrow keys. When done, select the OK character in
the bottom right corner to save the changed text. To exit without saving, press the Cancel
button.
12. Select menu item [6] Apply to view the calibration result. Verify the applied corrections in
the confirmation screen and press the Apply button to apply the adjustment to the probe,
or Cancel to exit without applying the adjustment.
13. Disconnect the probe from the HM40.
14. Connect the probe to the RFL100 Data Logger. See
Chapter 5 – Maintenance
Connecting the Probe (page
29).