Frequency Domain Using Frequency List - HP 8510 User Manual

Pulsed-rf network analyzer
Table of Contents

Advertisement

5. Select the maximum number of points required for the measurement, then perform the
appropriate measure- ment calibration.
Note that the HP 8510 pulse output is set to the active state (RF always On) during
measurement of the calibration standards. For external triggering, the pulse modulation is
operating during the calibration, so the pulse width or time span cannot be changed after
calibration.
6. For the next pulse pro le frequency repeat this sequence.
With several cal sets created in this way, you can select each pulse pro le frequency in turn by
recalling the corresponding cal set.

Frequency Domain Using Frequency List

In the above measurement calibration procedure it is necessary to perform a separate
calibration for each pulse pro le frequency. This is not a problem for simple response-only
calibrations, but when accuracy considerations require the use of 1-Port or 2-Port calibrations,
connecting the necessary sequence of standards repeatedly can be tedious. As an alternative,
the frequency list feature allows the standards to be connected once for all pulse pro le
frequencies.
Perform the calibration as follows.
1. Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY
4 DOMAIN 5
If the W annotation is not displayed, the wideband detector must be selected before
proceeding.
2. Press
NNNNNNNNNNNNNN
MORE
4 SYSTEM 5
3. Press
4 STIMULUS MENU 5
for each pulse pro le frequency to be measured. Each segment is de ned such that the
start and stop frequencies are identical and the number of points in each segment is the
number of points to be displayed in the pulse pro le measurement. For example, the
following sequence creates a frequency list to measure four pulse pro le frequencies, of 51
points each.
a. Press
NNNNNNNNNNN
ADD
4 START 5 4 3 5 4 G/n 5
b. Press
NNNNNNNNNNN
ADD
4 START 5 4 4 5 4 G/n 5 4 x1 5
c. Press
NNNNNNNNNNN
ADD
4 START 5 4 5 5 4 G/n 5 4 x1 5
d. Press
NNNNNNNNNNN
ADD
4 START 5 4 6 5 4 G/n 5 4 x1 5
e. Press
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE
FREQUENCY LIST
Figure 7-1 shows the resulting display. Proceed with the appropriate measurement
calibration.
7-2 Pulse Profile Domain Measurements
to select the frequency domain.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PULSE CONFIG
DETECTOR: WIDE BW
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
to display the Edit List menu. Create a segment
MORE
EDIT LIST
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER of POINTS
NNNNNNNNNNNNNN
.
DONE
NNNNNNNNNNNNNN
.
DONE
NNNNNNNNNNNNNN
.
DONE
. The frequency list will be measured.
.
NNNNNNNNNNNNNN
.
DONE
4 5 5 4 1 5 4 x1 5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8510a8510b8510c

Table of Contents