General Calibration and Measurement Sequence

Specify device input/output requirements.
Con gure test set for these levels, plus guardband.
Connect operating device and verify levels.
Adjust levels for best dynamic range.
Perform measurement calibration.
Measure operating device S-parameters.

General Calibration and Measurement Sequence Using Display Math

Specify device input/output requirements.
Con gure test set for these levels, plus guardband.
Connect operating device and verify levels.
Con gure setup for best dynamic range.
Connect standards and verify levels.
Change setup for best dynamic range during calibration.
Perform measurement calibration.
device.)
Connect appropriate standards and store reference traces.
Connect operating device.
Use normalization to view device parameters.
j
(Before saving cal set, con gure setup for operating
High Power Measurements 10-9