Maintenance

Maintenance Intervals

As required:
Replace the batteries, see page 11.
Annually:
Have the device calibrated at least once a year, as the measuring
inaccuracy increases gradually due to drift factors.
Only authorised persons may carry out the calibration.
We recommend commissioning Dräger for servicing the device.

Storage

The device consumes a small amount of current even when
switched off. Therefore, remove the batteries if the device will not
be used for a longer period of time.
If case of long-term storage:
Remove the batteries from the device if it will not be used for more
than six months. This prevents possible damage to the device
caused by leaking battery acid.

Cleaning

Do not dip the device into liquids and do not allow the connection
sockets to get wet!
For cleaning:
Wipe the device with a disposable cloth.
Do not use cleaning agents containing alcohol or solvents.