3.28 WAVeform Group

The WAVeform group deals with acquired waveform data.
:WAVeform?
Function
Queries all settings related to waveform data.
:WAVeform?
Syntax
:WAVeform:ACOunt?
Function
Queries the acquisition count.
:WAVeform:ACOunt?
Syntax
:WAVEFORM:ACOUNT? -> :WAVEFORM:
Example
ACOUNT 10
Description The function returns the acquired acquisition count.
The function returns "0" when data has not been
acquired at least once.
:WAVeform:BITS?
Function
Queries the bit length of the waveform data specified
by ":WAVeform:TRACe".
:WAVeform:BITS?
Syntax
:WAVEFORM:BITS? -> :WAVEFORM:BITS 16
Example
Description The bit length which is output with this command is
not the same as 'the valid bit length of A/D'. For the
SL1000, all analog Ch will be '16'.
:WAVeform:BYTeorder
Function
Sets or queries the byte order when using word
format of two bytes or more.
:WAVeform:BYTeorder {LSBFirst|MSBFirst}
Syntax
:WAVeform:BYTeorder?
:WAVEFORM:BYTEORDER LSBFIRST
Example
:WAVEFORM:BYTEORDER? -> :WAVEFORM:
BYTEORDER LSBFIRST
Description This command is valid for only the commands of the
WAVEfrom group.
3-64
:WAVeform:DIVision?
Function
Queries the Division value used when converting the
waveform data specified by 'WAVeform:TRACe' to
physical values.
:WAVeform:DIVision?
Syntax
:WAVEFORM:DIVISION? -> :WAVEFORM:
Example
DIVISION 24.000000E+03
Description The data is output per the current 'WAVeform:
FORMat' setting of the specified trace. The Division
value varies according to the kind of input module as
follows.
Input
Input
BYTE
Voltage
93.75
Tempearture
25.6
Strain
187.5
Acceleration
93.75
Frequency
93.75
RPM
93.75
RPS
93.75
Period
93.75
Duty
93.75
Power Freq
93.75
Pulses Width
93.75
Pulse Integ
93.75
Velocity
93.75
:WAVeform:END
Function
Sets or queries the final data point of the waveform
(main waveform) specified by 'WAVeform:TRACe'.
:WAVeform:END {<NRf>}
Syntax
:WAVeform:END?
<NRf>=0 to (total number of data points-1)
:WAVEFORM:END 100000
Example
:WAVEFORM:END? -> :WAVEFORM:END 100000
Description The total number of data points can be queried using
:WAVeform:CAPTure:LENGth?
:WAVeform:FORMat
Function
Sets or queries the format of the data to be
transmitted.
:WAVeform:FORMat {ASCii|BYTE|WORD}
Syntax
:WAVeform:FORMat?
:WAVEFORM:FORMAT WORD
Example
:WAVEFORM:FORMAT? -> :WAVEFORM:
FORMAT WORD
WORD
24000
0.1
48000
24000
24000
24000
24000
24000
24000
24000
24000
24000
24000
IM 720320-17E