Sanyo MDF-U333 Service Manual page 7

Biomedical freezer
Hide thumbs Also See for MDF-U333:
Table of Contents

Advertisement

MDF-U333
Items
Temperature controller
Temperature sensor
Temperature display
High/Low
Temperature
Alarm
Power failure
Remote alarm
contact
Control panel
Operation
panel
Rear
Key lock function
Self diagnosis function
Power switch
Compressor protection
Option
Microprocessor controlled system with non-volatilized memory.
Temp. set range : -18℃~-40℃
Thermistor
LED digital display (green colored) by 1℃.
Range : -50℃~50℃
Changeable SV±5℃~15℃. (Initial SV±10℃)
ALARM lamp brinks and buzzer sounds intermittently
(about 15 min. delay)
Remote alarm contact: turns ON during power failure alarm.
ALARM lamp brinks, buzzer sounds intermittently, remote alarm
contact turns ON during power failure alarm.
Remote alarm terminal 3P: contact capacity DC30v, Max2A, N.O.
–COM. (During power failure, temperature alarm: ON)
ALARM lamp
BUZZER: Buzzer key
ALARM TEST: Alarm test key
SET: To switch the SV and PV
>: To shift the set digit
∧: To change the SV
DEF: Defrost key
Power switch (Breaker switch)
Battery switch: Power failure alarm switch
Press > key for five seconds
L0: Unlock
L1: Lock
When each sensor is in fault,
Error code and internal temperature are displayed alternately.
Remote alarm contact turns ON with buzzer sounds.
Breaker switch
Overload relay
Chart recorder: MTR-G85
Mounting kit: MPR-S7(for bottom mounting)
: MDF-S740T(for top mounting)
− 4 −
Specification

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mdf-u537Mdf-u537d

Table of Contents