C I R C U I T D I A G R A; Mdf-U537/D Temperature Uniformity And Fluctuation - Sanyo MDF-U333 Service Manual

Biomedical freezer
Hide thumbs Also See for MDF-U333:
Table of Contents

Advertisement

MDF-U537 / D temperature uniformity and fuluctuation
Mains : 100V50Hz
: 35℃
AT
SV
:-20℃・-25℃・-30℃
No load Mesure with type T sensor
Measureing point : Center point of each container /chamber
MDF-U537
《 MDF-U537 》
Measuring point
1
2
3
4
5
6
7
8
Uniformity [deg.]  
Fluctuaation[℃] 
《 MDF-U537D 》
Measuring point
1
2
3
4
Uniformity [deg.]  
Fluctuaation[℃] 
5
6
7
8
9
Uniformity [deg.]  
Fluctuaation[℃] 
note)These data are reference only
SV -20℃
-19.7
-20.9
-21.1
-19.6
-19.9
-21.2
-18.7
-19.9
2.5
±2.0
SV -20℃
-17.7
-19.9
-18.8
-19.8
2.2
±3.1
-22.4
-21.3
-20.5
-17.9
-21.7
4.5
±3.2
− 48 −
MDF-U537D
SV -25℃
SV -30℃
-24.6
-25.7
-25.4
-24.7
-24.4
-25.1
-22.8
-24.1
2.6
±1.4
SV -25℃
SV -30℃
-22.4
-25.1
-24.9
-24.9
2.7
±2.1
-25.8
-25.3
-25.8
-23.3
-26.2
2.5
±2.6
-30.8
-31.7
-31.7
-29.4
-29.6
-30.7
-27.8
-29.7
3.9
±1.5
-27.4
-30.1
-30.8
-30.2
3.4
±0.9
-29.8
-29.5
-30.4
-28.1
-30.4
2.3
±1.3