Sanyo MDF-U333 Service Manual page 35

Biomedical freezer
Hide thumbs Also See for MDF-U333:
Table of Contents

Advertisement

F17 :Model code setting or checking
001 :MDF-U333
002 :MDF-U537
002 :MDF-U537D
Followings are initial setting values
Internal temp. SV : -30℃ (for U537D : both compartments)
Key lock mode : 0 (OFF)
High temperature alarm : SV+10℃(for U537D : both compartments)
Low temperature alarm : SV-10℃(for U537D : both compartments)
0 adjustment:+0℃(for temperature sensor)
Compressor start delay time : 1 minute
Communication ID : 000
Communication mode : 000
Alarm ring back time : 030 (30 minutes)
Set model code as MDF-U537, MDF-U537D to activate compressor protect function.
F21 : To set serial communication ID.
Input range: 000~255 (000 means that communication is OFF)
F22 : To set serial communication mode.
Control mode (3
Baud rate (2
st
1
digit is not used.
Note) Set control mode as "remote" unable to change SV and to activate defrosting.
F25 : To set value for alarm recovery (ring back) time.
Input value: 000,010,020,030,040,050,060 (In 000, recovery function is OFF)
11.
11.Differential (COMP. ON/OFF point on program)
11.
11.
COMP ON permit : Activate at +0.0℃ higher than SV
COMP OFF: Activate at -0.6℃ lower than SV
rd
0 : Local (initial value)
digit)
1 : Remote
nd
digit) 0 : 2400bps (initial value)
1 : 4800bps
2 : 9600bps
− 32 −