F08 :To adjust 0 of temperature sensor (setting range: -9.9℃~+9.9℃)
In F17: It is functioned in model code 002(MDF-U537), 003(MDF-U537D).
Return to PV display in model code 001(MDF-U333)
Setting method: In "F08" displayed, press SET key to display "00.0"(initial value).
Ex) When actual temperature of 1/2h chamber is shown –28.5℃and digital
display is shown -30℃, add +1.5 to value in F08.
Calculation = (-28.5-(-30)) = +1.5
"0"~"9" in first digit and first decimal place.
the first decimal place⇔the second digit.
F12 :Temperature of temperature sensor is displayed as per 0.1℃.
From second digit to first decimal point is shown in display.
It is not shown "-" in display when temperature is lower than -20℃.
(Ex. Actual temperature -35℃ → Display shows 35.0℃)
When temperature is within the range of -19.9℃~-0.1℃, "-" is displayed.
Range : -72.0~+83.0℃(±9.9℃)
F13 :Temperature of compressor protect sensor is displayed as per 1℃.
Range: -72~+163℃
F14 :Temperature sensor (solenoid valve control sensor for MDF-U537, lower compartment
temperature control sensor for MDF-U537D)
Temperature of each sensor is displayed as per 0.1℃
From second digit to first decimal point is shown in display.
It is not shown "-" in display when temperature is lower than -20℃.
(Ex. Actual temperature -35℃ → Display shows 35.0℃)
When temperature is within the range of -19.9℃~-0.1℃, "-" is displayed.
Range : -72.0~+83.0℃
Set temperature with
again to memorize set value. Input is done by differential value.
key enables to change only "0" and "-" in second digit and to change
key enables to shift the second digit ⇔ the first digit ⇔
− 31 −
key and
key and press SET key