W I R I N G D I A G R A; Mdf-U333 Temperature Uniformity And Fluctuation - Sanyo MDF-U333 Service Manual

Biomedical freezer
Hide thumbs Also See for MDF-U333:
Table of Contents

Advertisement

MDF-U333
Temperature uniformity and fluctuation
(Test condition)
Mains
: 1φ100V50/60Hz
AT
: 35℃
: -25℃・-20℃・-30℃
SV
No load
Measure with type T sensor
Measuring point:Center point of each container
(1).SV=-25℃
Measuring point
Top
3( 1/2 center air)
Middle
Bottom
Max./Min
Uniformity
Range(deg)
(2).SV=-20℃
Measuring point
Top
3( 1/2 center air)
Middle
Bottom
Max./Min
Uniformity
Range(deg)
(3).SV=-30℃
Measuring point
Top
3( 1/2 center air)
Middle
Bottom
Max./Min
Uniformity
Range(deg)
note)These data are reference only.
Temperature(℃)
1
2
4
5
Temperature(℃)
1
2
4
5
Temperature(℃)
1
2
4
5
− 42 −
Fluctuation(℃)
-21.8
-23.2
± 2.0
-24.3
-24.6
-24.2
-21.8/-24.6
2.8(±1.4)
Fluctuation(℃)
-17.1
-18.3
± 2.6
-19.1
-19.4
-19.0
-17.1/-19.4
2.3(±1.15)
Fluctuation(℃)
-26.3
-28.1
± 1.7
-29.3
-29.4
-29.1
-26.3/-29.4
3.2(±1.6)