Lumel NA5 Series User Manual

Lumel NA5 Series User Manual

One-channel digital-to-analogue meter with a multicoloured bargraph and a serial interface
Hide thumbs Also See for NA5 Series:

Advertisement

Quick Links

ONE-CHANNEL
DIGITAL-TO-ANALOGUE METER
WITH A MULTICOLOURED
BARGRAPH AND A SERIAL
INTERFACE
NA5 TYPE
USER'S GUIDE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NA5 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Lumel NA5 Series

  • Page 1 ONE-CHANNEL DIGITAL-TO-ANALOGUE METER WITH A MULTICOLOURED BARGRAPH AND A SERIAL INTERFACE NA5 TYPE USER’S GUIDE...
  • Page 3: Table Of Contents

    One-channel digital-to-analogue meter with multicoloured bargraph and a interface NA5 TYPE USER’S GUIDE CONTENTS Page 1. APPLICATION .....................5 2. SET OF THE METER ..................6 3. BASIC REQUIREMENTS, OPERATIONAL SAFETY ........7 4. INSTALLATION ....................9 5. SERVICING ....................13 6. RS-485 INTERFACE .................31 7. TECHNICAL DATA ..................49 8. BEFORE A FAILURE HAS BEEN DECLARED .........53 9. EXAMPLES OF NA5 METER PROGRAMMING ........55 10. ORDERING PROCEDURE ...............58 11. MAINTENANCE AND GUARANTEE .
  • Page 5: Application

    1. APPLICATION NA5 series meters with multicoloured bargraphs have an universal input desti- ned to measure temperature, resistance, voltage from shunts, standard signals, d.c. voltage and d.c. current. They can find application in various industrial fields, e.g. food industry, intermediate pumping stations, sewage treatment plants, chemical industry, weather stations, breweries. They are destined for the visualisation of the measured value and evaluation of change trends of checked technological processes. They can also find application in automation systems where programmed controllers are applied. Fig.1. View of the NA5 meter.
  • Page 6: Set Of The Meter

    NA5 meters can have in option: a continuous analogue output, a relay output, open collector (OC) type outputs and an RS-485 digital output. They are programmed by means of the keyboard and through RS-485. NA5 meters realise following functions: – measurement of the input quantity and displaying it on the display and the bargraphs, – recounting of the input signal into indication on the base of the individual linear characteristic, – arithmetical functions: raising to a power, extraction of roots, – programming of colours and bargraph resolutions, – signalling of alarm value setting exceedings, – recording of the measured signal in programmed time segments, – storage of maximal and minimal values, – programming of the measurement averaging time, – programming of the indication resolution, – deadlock of the parameter introduction by means of a password, – conversion of the measured quantity into a voltage or current output signal, – service of the RS-485 interface in MODBUS protocol, both in ASCII and RTU mode. 2. SET OF THE NA5 METER We deliver in the set: - NA5 meter 1 pc. - user’s guide 1 pc.
  • Page 7: Basic Requirements, Operational Safety

    3. BASIC REQUIREMENTS, SAFETY INFORMATION Symbols located in this service manual mean: WARNING! Warning of potential, hazardous situations. Especially important. One must acquaint with this before connecting the NA5 meter. The non-observance of notices marked by these symbols can occasion severe injuries of the personnel and the damage of the instrument. CAUTION! Designates a general useful note. If you observe it, handling of the me- ter is made easier. One must take note of this when the instrument is working inconsistently to the expectations. Possible consequences if disregarded ! In the security scope the meter meets the requirements of the EEC Low-Voltage directive (EN 61010 -1 issued by CENELEC).
  • Page 8 Ensure proper handling and avoid mechanical stress.  Do not bend any components and do not change any insulation distances.  Do not touch any electronic components and contacts.  Instruments may contain electrostatically sensitive components, which can easily  be damaged by inappropriate handling. Do not damage or destroy any electrical components since this might  endanger your health! 4. Electrical connection Before switching the meter on, one must check the correctness of connection  to the network. In case of the protection terminal connection with a separate lead one must ...
  • Page 9: Installation

    Before taking the meter housing out, one must turn the supply off.  The removal of the instrument housing during the guarantee contract period  may cause its cancellation. 4. INSTALLATION 4.1. Fitting Prepare a (44 x 137.5 ) mm hole in the panel. The thickness of the material +0.5 +0.5 from which the panel is made should be in the range 1...45 mm. The meter has screw terminal strips which enable the connection of 2.5 mm cross- section external conductors. Meter dimensions are shown on the fig. 2. max 92 Fig. 2. Meter overall dimension...
  • Page 10 4.2. External connection diagrams The description of terminal strips are shown on the fig. 3a. Connections of input signals are shown on the fig 3b and output signals on fig. 3c and 3d. The meter has programmable inputs. Maximal measuring ranges are given on figures. Power supply Supply of object 24 V d.c. transducers ...
  • Page 11 Resistance thermometer Resistance thermometer Thermocouple in two-wire system or in three-wire system or  300 mV voltage resistance measurement  600 V  40 mA  5 A  600 V voltage input  40 mA current input  5 A current input 24 V d.c. 24 V d.c.  40 mA  40 mA Three-wire object transducer Two-wire object transducer b/ Connection way of input signals...
  • Page 12 32 31 30 29 28 32 31 30 29 28 27 26 25 A1 A2 A3 A4 A5 A6 A7 A8 8 outputs of OC type 4 relay outputs c/ Connection way of digital and analogue output signals depending on the execution code IBM PC computer NA5 Meter IBM PC computer with an RS-485 card...
  • Page 13: Servicing

    5. SERVICING After connecting external signals and switching the meter on, its name and n100 also the current version of the program, e.g. . are displayed. After ca 3 seconds, the meter transits automatically into the working mode in which it carries out the measurement and the display of the measured value on the display and the bargraph. Depending on alarm parameter settings, the resolution and bargraph type, alarm thresholds are also displayed on the bargraph. The meter blanks automatically insignificant zeros. Key functions: 55-element bargraph (in three-colour execution) 29-element bargraph (in seven-colour execution) Upper alarm threshold Lower alarm threshold Display Acceptation key Escape key Value increasing key Fig. 4 Description of the NA5 frontal plate.
  • Page 14 - acceptation key - entry into the programming mode (hold down during ca 3 seconds), - entry into the chosen parameter level, - entry into the changing mode of the parameter value, - acceptation of the changed parameter value. - Key to increase the value - display of the minimal value (first pressure), maximal (second pressure), return to measurement (third pressure), - mowing on the preview menu or programming matrix, - change of the chosen parameter value - increasing of the value. - Escape key - entry into the menu of recording results, - entry into the preview menu or programming matrix, - exit from the preview menu or programming matrix, - escape from the parameter change. The pressure and hold down the key during 3 seconds causes the entry into the programming mode. The programming mode is protected by the safety code. The pressure and hold down the key during 3 seconds causes the entry into the preview menu and the menu of recorded values. One must move on the preview menu by means of the key. In this menu, only all programmed parameters except servicing parameters, are accessible to readout. The exit from the preview menu is operated by means of the key. It is also resl possible in the preview menu to review recorded values.
  • Page 15 view menu of recorder values. The recorded result number is displayed alternately n320/2174 with the value e.g. The moving on recorded values follows by means of the key. The pressure of this key longer than ca 2 seconds will cause the acceleration of the review. The pressure of the key in any moment will cause the lighting of the number of recorded results. The exit from the review of recorded values is operated by means of the key. The algorithm of the meter servicing is presented on the fig. 5. The appearance of the following symbols and inscriptions on the display means:...
  • Page 17 It is possible to change meter parameters: - from the meter keyboard ( p 5.1) Incorrectly introduced safety code Exceeding of the upper measuring range or lack of sensor Exceeding of the lower measuring range or short-circuited sensor Error of the conductor resistance compensation. No connected conductor or damaged conductor. - through RS-485 (p.6.) 5.1. Change of the NA5 meter parameters from the keyboard The pressure of the key during circa three seconds causes the display of the Inscription alternately with the set zero value by the manufacturer. The intro- duction of the correct code causes the entry into the programming mode. The fig.6 represents the transition matrix into the programming mode. One can move on groups of main parameters eg: Ch1, Ch2, bAr1, bAr2, Al1, Al2, etc, by means...
  • Page 18 Fig. 6. Transition matrix into the programming mode...
  • Page 19 Examples of value changing of the chosen parameter (parameter - symbol) 0000 000.0 00.00 0.000 Auto Example of value changing of the chosen parameter without changing of the decimal point (number parameter) Blanking of insignificant zeros Change of the Change of the Change of the Change of the flickering position...
  • Page 20 Programmable parameters of NA5 meter Table 1 Symbol on the Range of changes Parameter description display Reesistance thermometers: Kind of input - Pt100 - Pt500 pt10 - Pt1000 Thermocouples: te-j - thermocouple, type J te-h - thermocouple, type K te-n - thermocouple, type N te-e - thermocouple, type E te-r - thermocouple, type R te-s - thermocouple, type S te-t - thermocouple, type T - resistance to 10 k napl - voltage to  300 mV napH - voltage to  600 V nnal - current to  40 mA nnaH...
  • Page 21 auto Kind of compensation of sensor - automatic compensation working conditions changes: (in case of resistance thermometers and - In case of a resistance thermometer resistance measurement it requires a and resistance measurement, it 3-wire line.) concerns the compensation of 0.0...60.0C - value of the reference the resistance changes of the temperature for thermocouples. conductor linking the sensor with 0.0...40.0 - resistance of two conduc- the meter, tors for resistance thermometers and...
  • Page 22 typb Bargraph type 0neC - „ one colour „ bargraph, :ntr - „ interval „ bargraph, sect - „ sector” bargraph, p:nt - „ point bargraph, tren -”trend” bargraph. Fig. 10 explains bargraph types. . colr Bargraph colour - bargraph switched off, - red, - green, - red + green other colours are accessible only in meters with a 7-colour bargraph. - blue, - red + blue, - green + blue, - red + green + blue. Parameter to set the „magnifier” Setting possibility: -1999... 9999 on the bargraph. Lower threshold. Value on the display at which the bargraph is to be blank.
  • Page 23 Lower alarm threshold Setting possibility: - 1999... 9999 Upper alarm threshold Setting possibility: - 1999... 9999 typa - normal, Alarm type Fig. 8 shows alarm types - switched on, - switched off, H_0n - manually switched on. Till the time of the alarm type change, the alarm output is being permanently switched on. H_0f - manually switched off. Till the time of the alarm type change, the alarm output is being permanently switched off. Delay of the alarm operation. The 0.0... 999.9 parameter is defined in seconds, i.e. one must give the time in seconds Introduction of 0.0 causes the ope- after which the alarm will operate after ration at the moment of the alarm...
  • Page 24 Curl Colour of the lower threshold - alarm marker switched off. alarm marker. - red, CurH Colour of the upper threshold - green, alarm marker. - red + green, Other colours are accessible only in meters with a 7-colour bargraph. - blue, - red + blue, - green + blue, - red + green + blue, Fig. 10 explains the idea of CurL and CurH parameters :nd0 Switching off or on of the individual - characteristic switched on,...
  • Page 25 2400 baud Baud rate of the RS-485 interface. - 2400 b/s 4800 - 4800 b/s 9600 - 9600 b/s tryb Kind of transmission through the - interface switched off RS-485 interface. a8n1 - ASCII 8N1 a7e1 - ASCII 7E1 a7o1 - ASCII 7O1 r8n2 - RTU 8N2 r8e1 - RTU 8E1 r8o1 - RTU 8O1 r8n1 - RTU 8N1 Device address Setting possibility: 0...247 Manufacturer’s parameters. The pressure of the key causes Manufacturer’s parameters are...
  • Page 26 Switching the recording on or - recording switched off off. At the moment of switching the - recording switched on recording on, the meter erases the previous stored values. go_r Hour of recording start Setting possibility: 00:00:00 ... 23:59:59 Time format: hh:mm:ss date Date of recording start Setting possibility: 70.01.01 ... 38.12.31 Date format: yy.mm.dd It is an information parameter. It not serves to define the date from which the recording is to begin, but only to inform when the recording 1nte...
  • Page 27 a) nor (PrH > PrL) 1 - Relay switched on 0 - Relay switched off Measured value b) nor (PrH < PrL) 1 - Relay switched on 0 - Relay switched off Measured value c) OFF 1 - Relay switched on 0 - Relay switched off Measured value d) On...
  • Page 28 Displayed quantity Any characteristic slope d_Y2 d_Y1 Any displacement of the characteristic I_H1 I_H2 Measured quantity Value I_H1 on the meter input => value d_Y1 on the display. Value I_H2 on the meter input => value d_Y2 on the display other characteristic points are calculated Quantity on the analog output Any characteristic slope O_Y2 O_Y1 Any displacement of the characteristic d_H1 d_H2 Displayed value Value d_H1 on the display => value O_Y1 on the analogue output. Value I_H2 on the display => value O_Y2 on the analogue output other characteristic points are calculated Fig.
  • Page 29 Examplary settings of the bargraph and the alarm, ex. 1 Type of Curl = G (green) Notes bargraph Curl CurH = r (red) = rG (red+green) 0neC Value under the value prl :ntr Value between Value ower sect p:nt Value without changes in time tren Value...
  • Page 30 – The meter currently checks the value of the introduced parameter at the mo- ment. In case when the introduced value exceeds the upper or the lower range of changes given in the table 1, the meter will not make the parameter record. – In case of the Input type change, a simultaneous change of the decimal point follows, optimally for the given input. – After the supply decay, the current time is reset. – The recording switching off follows in following cases: switching the recording off from the programming matrix, change of the input type, change of the recording time start or the recording time interval, setting Cnt=0, filling of the memory, and at a renewed switching of the meter on to the network. – In case of a Intr or Sect bargraph type, it is possible to set only one CurL and Curh markers (from one alarm). The others are automatically erased. – Max and Min values are erased in case of changes: input type, individual cha- racteristic (on, off), writing of standard parameters. 6. RS-485 INTERFACE Standard parameters of the NA5 meter Table 2 Parameter description Standard value Parameter description Standard value - 20.00 nnal (± 40 mA) 20.00 lo:n - 20.0 typa Hi:n 20.00 func...
  • Page 31: Rs-485 Interface

    date 70:01:01 :nte 00:15:00 Hour 00:00:00 gor_r 00:00:00 NA5 programmable digital meters have a serial link of RS-485 standard to communicate in computer systems and with other devices fulfilling the master function. The MODBUS asynchronous character communication protocol has been implemented on the serial link. The transmission protocol describes information exchange procedures between devices through the serial link.
  • Page 32 Company specification. Set of serial link parameters of meters in the MODBUS protocol: meter address - 1... 247  baud rate - 2400, 4800, 9600 bit/s  working mode - ASCII, RTU  information unit - ASCII: 8N1, 7E1, 7O1  - RTU: 8N2, 8N1, 8E1, 8O1 maximal response time 500 ms  The configuration of serial link parameters is described in the further part of the user’s manual. It consists on the establishment of the baud rate (bAud parameter), device address (Adr parameter) and the type of information unit (trYb parameter). Note: Each meter connected to the communication network must have: a unique address, different from addresses of other devices connected  in the network, an identical baud rate and information type.  6.3. Description of the MODBUS protocol functions Following functions of the MODBUS protocol have been implemented in NA5 meters: Function description...
  • Page 33 Read-out of n-registers (code 03 h) Function is inaccessible in the publication mode. Example: readout of 2 registers beginning from the register with the address 1 DBDh (7613) in RTU mode. Request: Register Register Number of Number of Check- Device address address registers registers Function address 52 43 Response: Check- Device Number Value from the register Value from the register Function address of bytes 1DBD (7613) 1DBE (7614)
  • Page 34 Response: Register Register Number of Number of Check- Device Function address address registers registers address (CRC) D7 80 Report identifying devices (code 11h) in RTU mode Example: Data readout identifying the device for NA5 with a universal input. Request: Device Checksum Function address (CRC) C0 2C Response: Device Number Device Device Field depending on Check- Function address...
  • Page 35 Example: Work in RTU mode: e.g. Mode = RTU 8N2 (value 0x02 in case of readout/record through the interface). NA5 meter Execution with a voltage analogue output: 00, No. of the software version: 1.00, Device address set on: Adr = 0 x 01, For such a meter the frame has the following form: Check- Device Number Device Device Field depending on the Function address of bytes identifier state device type (CRC) 00 00 3F 80 00 00 FE D7 6.4. Register map of NA5 meters Register map of NA5 meters Table 4.
  • Page 36 6.5. Registers for recording and readout NA5 meter Symbol Range Description Identifier 7200 7600 Device identifier Value Channel 7202 7601 0...1 No occurs number input 7204 7602 0...14 Channel input type <Channel number> Value Pt100 RTD Pt500 RTD Pt1000 RTD J thermocouple K thermocouple N thermocouple E thermocouple...
  • Page 37 LoIn Lower value of the input range -1999... 9999 7206 7603 Upper value of the input range HiIn 7208 7604 -1999... 9999 Function 7210 7605 0... 7 Arythmetical function Value Switched off Squaring Extraction of roots Compensation of the conductor resistan- Compens.
  • Page 38 No occurs 7228 7614 Bargraph Bargraph type 7230 7615 0... 4 type Value One-colour (OnEC) Change of colour after exceeding the alarm threshold (the colour change the whole bargraph) (Intr) Change of colour after exceeding the alarm threshold (Three-segment change of colour) (SEct) One-colour bargraph, alarm markers in another colour...
  • Page 39 7240 7620 No occurs Value Channel 1 Channel 2 7242 7621 -1999... 9999 Alarm lower threshold <Alarm No> 7244 7622 -1999... 9999 Alarm upper threshold <Alarm No> Type a 7246 7623 0... 4 Alarm type <Alarm No> Value Normal Switched on Switched off Manually switched on Manually switched off...
  • Page 40 Bargraph colour after exceeding the CURH 7254 7627 0... 7 upper alarm threshold <Alarm No> Value Bargraph switched off (OFF) Red (r) Green (G) Red + Green (rG) Other values accessible only in meters with RGB diodes Blue (b) Red + Blue (rb) Green + blue (Gb) Red + Green + Blue (rGb) 7256...
  • Page 41 RTU 8E2 RTU 8O2 RTU 8N1 Address 7272 7636 0... 247 Choice of the device address Test 7274 7637 0... 1 Test of the display Value Lack of operation Test Hour 7276 7638 0... 23.5959 Current time This parameter occurs with four places after the decimal point in format gg,mmss, where: gg - means hours,...
  • Page 42 7288 7644 1... 31 Day of the recording start Year, Month, Day are information parameters (they do not serve to define the date from which the recording is to be start). Erasing of 7290 7645 0... 1 Erasing of the minimal value minimum Value Lack of operation...
  • Page 43 Time of the memorised 7326 7663 0... 23.5959 Time of memorised value in memory value This parameter occurs with four places after the decimal point in format gg,mmss, where: gg - means hours, mm - means minutes, ss - means seconds In case when introducing and incorrect time, the meter will correct it automa- tically.
  • Page 44 Number of the memori- 7332 7666 0... 750 Number of memorised value in memory, sed value placed in the first register of the buffer Value Memory is empty 1... 750 Number of the memorised value Number of 7334 7667 0... 750 Number of recorded buffer registers recorded registers...
  • Page 45 6.6. Registers only for readout Name Unit Quantity name Identifier 7000 7500 Constant identifying the device Register describing the current state of Status 1 7002 7501 the meter Register describing the current state of Status 2 7004 7502 the meter It is the register defining the control Steering 7006...
  • Page 46 Description of the Status 1 register Position of the decimal point X X 15 14 13 12 11 10 9 bits MSB Bit-15... 9 No used Bit-8 Error of the conductor resistance compensation 0 - lack of error 1 - signalling of the compensation error Bit-7... 5 Position of the decimal point 000 - lack 001 - 000.0 010 - 00.00...
  • Page 47 Bit-2 Individual characteristic 0 - individual characteristic switched off 1 - individual characteristic switched on Bit-1...0 Kind of output (voltage, current) 00 - lack of analogue output 01 - current 10 - voltage Description of the status 2 register Working mode and information unit X X 15 14 13 12 11 10 9 bits MSB Bit-15. No used Bit-14...13 Record of measurement results in memory 0 - Recording switched off 1 - Recording switched on Bit-12 State of alarm 8...
  • Page 48 Bit-10 State of alarm 6 0 - off 1 - on Bit-9 State of alarm 5 0 - off 1 - on Bit-8 State of alarm 4 0 - off 1 - on Bit-7. State of alarme 3 0 - off 1 - on Bit-6 State of alarm 2 0 - off 1 - on Bit-5 State of alarm 1 0 - off 1 - on Bit-4...2 Working mode and information unit...
  • Page 49: Technical Data

    7. TECHNICAL DATA INPUTS: Pt100 (- 200... + 850)C Pt500 (- 200... + 850)C Pt1000 (- 200... + 850)C J (Fe-CuNi) (- 100... + 1100)C K (NiCr-NiAl) (-100... + 1370)C N (NiCrSi-NiSi) (- 100... + 1300)C E (NiCr-CuNi) (- 100... + 850)C R (PtRh13-Pt) (0... + 1760)C S (PtRh10-Pt) (0... + 1760)C T (Cu-CuNi) (- 50... + 400)C Resistance measurement 0... 10 k  300 mV, Voltage measurement input resistance > 9 M,  600 V, Voltage measurement input resistance > 4.2 M  40 mA, Current measurement input resistance < 4  Current measurement 0...  5 A, input resistance = 10 m  10% Measuring subranges (preserving the class): Pt100 320C Pt500...
  • Page 50 OUTPUTS: – Analogue outputs galvanically isolated, with a resolution = 0,025% of the range - current programmable: 0/4...20 mA load resistance  500  - or voltage programmable: 0...10 V load resistance  500  - output response time 100 ms - output error 0.2% of the range - additional error due to ambient  (0.1% of the range/10K) temperature changes: – Relay output 4 relays; voltageless make contacts - maximal load: voltage 250 V a.c., 150 V d.c., current 5 A 30 V d.c., 250 V a.c., resistance load 1250 VA, 150 W. Programmable alarm thresholds; Three types of alarms; Hysteresis defined by means of the lower and upper alarm threshold; Signalling of alarm operation on the bargraph; – 8 outputs of open collector (OC) type voltageless, OC type with npn transistor (max. load 25 mA) range of connected voltage: 5...30 V d.c.
  • Page 51 0.1% of measuring range 1 digit Basic error: 0.2% of measuring range 1 digit (for thermocouples R, S, T) Additional errors in nominal working conditions when measuring the temperature: - compensation of reference junction  1C temperature changes - compensation of conductor  0.1% of the range resistance changes  (0.05% of the range/10K) - from ambient temperature changes Averaging time min 200 ms min 500 ms (temperature ranges) Rated operation conditions: - supply voltage depending on the execution code 95...230...253 V a.c./d.c. 20...24...40 V a.c./d.c. - supply a.c. voltage frequency 40...50/60...440 Hz - ambient temperature - 10...23...55C - storage temperature - 25...+85C - relative humidity...
  • Page 52 Readout field (depending on execution): 4 LED seven-segment LED display, character height: 7 mm indication range: -1999...9999 bargraph length: 88 mm - 55 segments in three-colour execution - 29 segments in seven-colour execution Bargraph resolution programmable Bargraph accuracy ± 0.5 segment Servicing three keys: Ensured protection degree:...
  • Page 53: Before A Failure Has Been Declared

    8. BEFORE A FAILURE WILL BE DECLARED In case of incorrect symptoms please to acquaint with the table below. SYMPTOMS PROCEDURE 1. Lack of indications on the display. The Check the connection of the feeder cable. bargraph indicates nothing. 2. The time is displayed on the display, The number of measurements Cnt = 0 has been e.g.
  • Page 54 8. A result inconsistent with our expecta- Check if the individual characteristic is not switched tions appears on the display. on. In case of necessity enter into the programming matrix and accept the SEt parameter. The meter will introduce parameters in accordance with the table 2. 9.
  • Page 55: Examples Of Na5 Meter Programming

    9. EXAMPLES OF NA5 METER PROGRAMMING Example 1. Programming of the individual characteristic. If we want to programme so that to the value 4.00 mA will correspond the value 0 on the display, whereas the value 100, to the value 20.00 mA, one must: enter into the programming mode and choose the D_ P parameter  responsible for the decimal point. Set the decimal point on 00000 choose the Ind parameter and switch the individual characteristic On  choose the I_H1 parameter and introduce the value 4.00  transit on the d_Y1 parameter and introduce the value 0  transit on the I_H2 parameter and introduce the value 20.00  transit on the d_Y2 parameter and introduce the value 100  Example 2 Programming of an inverse individual characteristic. If we want to programme so that to the value 4.00 mA will correspond the value 120.5 on the display, and the value 10.8, to the value 20.00 mA, one must: enter into the programming mode and choose the D_P parameter responsible ...
  • Page 56 Example 4 Programming of an alarm operating in a set interval with delay. If we want that the alarm 1 will be switched on in the interval from 100 V to 300 V and operate only after 10 seconds, one must: enter into the programming mode, choose the PrL parameter of the alarm 1  and introduce the value 100 transit on the PrH parameter of the alarm 1 and introduce the value 300  transit on the tYPA parameter of the alarm 1 and select the function On  transit on the dLY parameter of the alarm 1 and introduce the value 10.0  in case of the alarm state duration for a time longer than 10 seconds, the meter will switch the alarm relay on Example 5 Programming of an analog output If we want to programme so that to the displayed value 0.00 mA will correspond the value 4.00 on the analogue output, whereas to the value 20.00 mA , the value 20.00 mA, one must: enter into the programming mode, choose the IndO parameter and switch the  individual characteristic On choose the d_H1 parameter and introduce the value 0.00 ...
  • Page 57 enter into the programming mode, choose the Go_r parameter and introduce  the value 12:30, transit into IntE parameter and introduce the value 00:00:20,  choose the rEc parameter and switch the recording in On,  after exiting from the programming matrix, the memory will be erased and  the meter will begin recording results evry 20 sec, from 12:30. After filling the memory, the recordingwill be switching off.
  • Page 58: Ordering Procedure

    10. ORDERING PROCEDURE Table 5 NA5 METER WITH BARGRAPH XX X Bargraph colour: 3 colours (R, G, R + G) ........T 7 colours (R, G, B, R+G, R+B, G+B, R+G+B) ...M Display colour: without display* ............0 red ............R green ................ G blue ................
  • Page 59 Example of NA5 ordering: Code: NA5 M G U 1 1 4 1 0 00 0 means: M - NA5 meter with two 7-colour bargraphs, G - digital LED displays of green colour, U - universal input (table 6), 1 - current analogue output signal 0/4...20 mA, 1 - RS-485 digital output signal, 4 - additional outputs consisting of 4 relays, 1 - supply: 95...253 V a.c./d.c., 0 - socket-screw plug terminals, 00 - standard execution, 0 - without a quality inspection certificate. In case of a custom-made execution or if you need some more additional technical information, please write to or phone our Export Department. Input signals Table 6 Universal Resistance thermometer: input Pt100 (- 200... +850)C Pt500 (- 200... +850)C Pt1000 (- 200... +850)C Thermocouple: J (Fe-CuNi) (- 100... +1100)C...
  • Page 60: Maintenance And Guarantee

    11. MAINTENANCE AND GUARANTEE The NA5 meter does not require any periodical maintenance. In case of some incorrect unit operations: 1. From the shipping date, during the period given in the annexed guaran- tee card One should take the meter down from the installation and return it to the Manufacturer’s Quality Control Dept. If the unit has been used in compliance with the instructions, the Manufacturer guarantees to repair it free of charge. 2. After the guarantee period: One should turn over the meter to repair in a certified service workshop. The disassembling of the housing causes the cancellation of the granted guaran- tee. Spare parts are available for the period of five years from the date of purchase. The Manufacturer policy is one of continuous improvement and we reserve the right to make changes in design and specification of any products as engineering advances or necessity requires and revise the above specification without notice.
  • Page 64  CHART AND PAPERLESS RECORDERS  POWER CONTROL UNITS and INVERTERS  AUTOMOTIVE DASHBOARD INDICATORS  ACCESSORIES FOR MEASURING INSTRUMENTS  MEASURING SYSTEMS (ENERGY, HEAT, CONTROL)  CUSTOM-MADE PRODUCTS WE ALSO OFFER OUR SERVICES IN THE PRODUCTION OF:  ALUMINIUM ALLOY PRESSURE CASTINGS  PRECISION ENGINEERING AND THERMOPLASTICS PARTS QUALITY PROCEDURES: According to ISO 9001 and ISO 14001 international requirements. All our instruments have CE mark. For more information, please write to or phone our Export Department. Lubuskie Zak³ady Aparatów Elektrycznych LUMEL S.A. ul. Sulechowska 1, 65-022 Zielona Góra, Poland Tel.: (48-68) 329 51 00 (exchange) Fax: (48-68) 329 51 01 e-mail:lumel@lumel.com.pl http://www.lumel.com.pl Export Department: Tel.: (48-68) 329 53 02 or 53 04 Fax: (48-68) 325 40 91 e-mail: export@lumel.com.pl...

Table of Contents