Possible Problems and resolutions

Problems
SpO
or PR
1.Finger is not plugged correctly
2
can not be
2.Patient's Oxyhemoglobin value is
shown normally
too low to be measured.
1.Finger might not be plugged
SpO
or PR is
deeply enough.
2
shown unstably
2.Finger is trembling or patient's
body is in movement status.
1.Power of batteries might be
The Oximeter
inadequate or not be there at all.
can not be
2.Batteries might be installed
powered on
incorrectly.
3.The Oximeter might be damaged.
1.The product is automatically
Indication
powered off when no signal is
lamps are
detected longer than 8 seconds
suddenly off
2. Lower power
"Error3"or
1.Low power
"Error4" is
2.Mechanical malfunction for
displayed on
receive-emission tube
screen.
3. Amp circuit malfunction.
Possible reason
- 10 -
Solution
1. Retry by plugging the finger.
2. Try some more times, If you
can make sure about no problem
existing in the product. Please go
to a hospital for exact diagnosis.
1.Retry by plugging the finger.
2.Try not to move.
1.Please replace batteries
2.Please reinstall the batteries
3.Please contact with local customer
service centre.
1.Normal
2.Replace the batteries
1.Change batteries.
2.Please contact the local customer
service center.
3.Please contact the local customer
service center.