Diagnostics Questions And Answers - Cardiac Science SURVIVALINK AED Operation & Service Manual

Automated external defibrillator
Table of Contents

Advertisement

Section 6: Maintenance & Troubleshooting
10.
Q
My Survivalink AED is sounding an audible alert. Why? How do I stop it?
A
The audible alert indicates that the self-test detected a need for maintenance or cor-
rective action. Determine the maintenance required by using the Troubleshooting
Table in this chapter.
Opening and closing the lid may turn OFF the audible alert until the next self-test.
The" Status" indicator, however, will remain RED.
11.
Q
How long will the audible alert sound before the battery is worn down?
A
A full, Extended Life battery can continue an audible alert for approximately one
year.
12.
Q
When I open the lid, why do I get the voice prompt "Data in memory. Do not push
'Resume' button until data is downloaded, unless you must do a rescue?" How do I
get the message to stop occurring?
P
age 62

Diagnostics Questions and Answers

The following table answers frequently asked questions about
Diagnostics:
Questions and Answers
©2002 Cardiac Science, Inc.