Aimer Adjustment - Siemens sitrans LR560 Quick Start Manual

Hide thumbs Also See for sitrans LR560:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Environment
• Provide easy access for viewing the display and programming via the hand programmer.
• Provide an environment suitable to the housing rating and materials of construction.
Sun Shield
The LR560 display can be protected by an
optional sun shield if the instrument will be
mounted in direct sunlight

Aimer Adjustment

Aiming is not
required for signal
optimization with
78 GHz frequency.
3" flange
1)
For 4" and 6" Aimer: Loosen the set screws in the locking ring.
Holding the electronics enclosure firmly, loosen the Aimer locking ring using the supplied
C spanner, until the LR560 drops down slightly. The enclosure can then be turned freely.
2)
Direct SITRANS LR560 so the antenna is pointed at an angle perpendicular to the
material surface, if possible.
7ML19985XF82
Aiming will assist
in measuring
material in the
cone.
4 and 6" flange
SITRANS LR560 (HART)
Aimer
Page EN-7