Section 2 - Installation
2.1 Unpacking
Remove the packing list and verify that you have received all your equipment.
If you have any questions about the shipment, please call our Customer Service
Department.
We can also be reached on the Internet.
When you receive the shipment, inspect the container and equipment for any
signs of damage. Note any evidence of rough handling in transit. Immediately
report any damage to the shipping agent.
The carrier will not honor any damage claims unless all shipping material is saved
for inspection. After examining and removing contents, save packing material and
carton in the event reshipment is necessary.
The following items are supplied in the box:
• BB701 Blackbody Calibration Source
• User's Manual
• Calibration Certificate
• Power Cord
• Connector

2.2 Mounting

Mount the unit on a bench, table top or shelf in a horizontal position and operate
at least ten inches from any air obstructions to the fan, front panel, rear panel,
and top of the unit. Operate the unit in an ambient environment between the
specified 4.4 to 43°C (40 to 110°F).

2.3 Ambient Temperature

The target plate of the BB701 can achieve any temperature within the specified
temperature range of -18 to 149°C (+0 to 300°F) when being operated in normal
ambient temperature environments. The maximum specified target plate
temperature of 149°C (300°F) can be achieved over the entire specified ambient
temperature range. However, the lower target plate temperatures are more
difficult to attain at increased ambient temperatures. As long as the ambient
temperature does not exceed 29.4°C (85°F), the target plate will achieve its lower
limit temperature of -18°C (0°F). The minimum target plate temperature the unit
can achieve is proportionally worse with increased ambient temperature.
An increase of 1°C is accompanied by an increase in minimum target plate
temperature of 0.3°C (An increase of 1°F is accompanied by an increase in
minimum target plate temperature of 0.5°F).
NOTE:
2
BB701
2-1