    Did you know? Replacing your mask and supplies is one of the most important steps in maintaining effective therapy. Replace when... • The cushion is no longer clear and transparent. • The cushion is cracked, torn or damaged in any way. • The headgear is stretched and needs to be tightened more than when it was new.
  • Page 3 Quattro™ FX Quattro™ FX for Her FULL FACE MASK / MASQUE FACIAL / MASCARILLA FACIAL / MÁSCARA FACIAL Fitting / Mise en place / Colocación / Colocação e ajuste...
  • Page 4 Headgear position / Position du harnais / Posición del arnés / Posição do arnês Cushion check / Vérification de la bulle / Revision de la almohadilla / Verificação da almofada Disassembly / Démontage / Desmontaje / Desmontagem...
  • Page 5 Reassembly / Remontage / Montaje / Remontagem...
  • Page 6 Quattro FX Quattro FX for Her...
  • Page 7 Part number / Code produit / Article / Descripción / Descrição Número de pieza / Código do Artículo / produto Item Quattro FX Quattro FX for Her Spring frame / Entourage flexible / Armazón elástico / Armação — — flexível Vent / Orifice de ventilation / —...
  • Page 8 Valve clip / Clip de la valve / Broche de la válvula / Presilha da — — válvula Valve & clip / Valve et clip / Válvula — — y broche / Válvula e presilha Complete system / Système 61700 (S) 62501 (S) complet / Sistema completo / 61701 (M)
  • Page 9 Quattro FX for Her ™ FULL FACE MASK Thank you for choosing the Quattro FX/Quattro FX for Her. Please read the entire manual before use. Intended use The Quattro FX/Quattro FX for Her channels airflow noninvasively to a patient from a continuous positive airway pressure (CPAP) or bilevel device.
  • Page 10 Ensure the soft sleeves are fitted at all times during use. Using your mask When using your mask with ResMed CPAP or bilevel devices that have mask setting options, refer to the Technical specifications section in this user guide for mask selection options.
  • Page 11 Ensure your headgear is positioned correctly. Refer to the Headgear check illustrations in this user guide. • The Quattro FX/Quattro FX for Her cushion is designed to sit mid-bridge. Refer to the Cushion check illustrations for correct positioning. Removing your mask To remove your mask, disconnect the lower headgear clips and pull the mask and headgear up over your head.
  • Page 12: Disassembly Steps

    Disassembly Before disassembling your mask, detach the air tubing and unclip the lower headgear clips from the spring frame. Refer to the disassembly illustrations in this guide. Disassembly steps Remove the elbow assembly from the mask. Gently pull the spring frame away from the frame. Remove the soft sleeves from the spring frame.
  • Page 13 Cleaning your mask in the home Your mask and headgear should only be gently handwashed. Daily/After each use: • Disassemble the mask components according to the disassembly instructions. • Thoroughly clean the separated mask components (excluding headgear), by gently rubbing in warm (approximately 86°F/30°C) water using mild soap.
  • Page 14: Troubleshooting

    Reprocessing the mask between patients This mask should be reprocessed when used between patients. Cleaning, disinfection and sterilization instructions are available from the ResMed website, www.resmed.com/masks/sterilization/americas. If you do not have internet access, please contact your ResMed representative. Troubleshooting Problem/possible cause Solution...
  • Page 15 Talk to your clinician to have your face size checked size. against fitting template. Note that sizing across different masks is not always the same. The Quattro FX/Quattro FX for Her cushion is designed to sit mid-bridge. Spring frame incorrectly Remove the spring frame from the frame, then fitted to frame.
  • Page 16 WARNING • The technical specifications of the mask are provided for your clinician to check that they are compatible with the CPAP or bilevel device. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect the CPAP or bilevel device function.
  • Page 17: Technical Specifications

    Technical specifications Pressure–flow The mask contains passive venting Pressure Flow curve to protect against rebreathing. As a (cm H (L/min) result of manufacturing variations, the vent flow rate may vary. Mask pressure (cm H Dead space Physical dead space is the empty volume of the mask to the end of the swivel.
  • Page 18 Notes: • The mask system does not contain PVC, DEHP or phthalates. • This product does not contain natural rubber latex. • The manufacturer reserves the right to change these specifications without notice. Storage Ensure that the mask is thoroughly clean and dry before storing it for any length of time.
  • Page 19: Limited Warranty

    90 days or in the case of disposable masks the maximum use period. This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use ResMed will repair or replace, at its option, the defective product or any of its components.
  • Page 20 Manufacturer: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia See www.resmed.com for other ResMed locations worldwide. Protected by patents: AU 777033, AU 785376, EP 1187649, EP 1314445, EP 1582230, EP 1640034, HK 1057714, JP 4354664, JP 4607913, NZ 538699, US 6439230, US 6581594, US 6823865, US 7159587, US 7926487, US 7669599.
  • Page 21 618179/2 2017-09 QUATTRO FX QUATTRO FX FOR HER USER amer ENG 618179 Global leaders in sleep and respiratory medicine www.resmed.com...

Table of Contents