E. The display mode can be changed
F. Screen brightness can be changed
G. A pulse sound indication
H. With alarm function
I I I . With SpO
J. It can be connected with an external oximeter probe
K K K . Real-time data can be transmitted to computers
L. With clock function
4 4 4 .2 .2 .2 Main
A. A. A. Measurement
Measuring range: 0%~100%
Accuracy:
When the SpO
below 70% unspecified
B. B. B. Measurement
Measuring range:25bpm~250bpm
Accuracy: ±2 bpm or ±2% (select larger)
C. C. C. Resolution
SpO
D. D. D. Measurement
SpO
pulse rate error is ±2 bpm or ±2% (select larger).
E. E. E. Resistance
The deviation between the value measured in the condition of man-made light or indoor natural light
and that of darkroom is less than ±1%.
F. F. F. Power
G. G. G. Optical
Red light (wavelength is 660nm,6.65mW)
Infrared (wavelength is 880nm, 6.75mW)
Downloaded from
www.Manualslib.com
value and pulse rate value of storage, the stored data can be uploaded to computers
2
Main

Main Parameters

Parameters
Parameters
Measurement of of of SpO
Measurement
2
measuring range is 70%~100%,the permission of absolute error is ±2%;
2
Measurement
Measurement of of of pulse
pulse
pulse rate
rate
rate
Resolution
Resolution
: 1%, Pulse rate: 1bpm.
2
Measurement
Measurement Performance
Performance
Performance in in in Weak
and pulse rate can be shown correctly when pulse-filling ratio is 0.4%. SpO
2
Resistance
Resistance to to to surrounding
surrounding
light
surrounding light
light
Power supply
Power
supply
supply requirement:
requirement:
requirement: : 3.6 V DC ~ 4.2V DC.
Optical
Sensor
Optical Sensor
Sensor