Rosemount 8712 Reference Manual

Rosemount 8712 Reference Manual

Remote mount magnetic flowmeter system
Table of Contents

Advertisement

Reference Manual
00809-0100-4664, Rev BA
January 2010
Rosemount 8712
Remote Mount Magnetic Flowmeter System
www.rosemount.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 8712 and is the answer not in the manual?

Questions and answers

Oscar Ramirez
May 27, 2025

Buenos días, realicé comprobación de parámetros en un equipo rosemount 8712 , y al realizar el auto zero trabaja pero parpadea la pantalla y dice auto zero fallido. Les agradezco su apoyo.

Subscribe to Our Youtube Channel

Summary of Contents for Rosemount 8712

  • Page 1 Reference Manual 00809-0100-4664, Rev BA January 2010 Rosemount 8712 Remote Mount Magnetic Flowmeter System www.rosemount.com...
  • Page 3 The products described in this document are NOT designed for nuclear-qualified applications. Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate readings. For information on Rosemount nuclear-qualified products, contact your local Emerson Process Management Sales Representative. www.rosemount.com...
  • Page 5: Table Of Contents

    Connections ........2-13 Rosemount Sensors ........2-13 Transmitter to Sensor Wiring.
  • Page 6 Reference Manual 00809-0100-4664, Rev BA Rosemount 8732 January 2010 Basic Setup..........3-8 Tag .
  • Page 7 Diagnostics Licensing the 8712 Diagnostics......C-2 Tunable Empty Pipe Detection ......C-2 Tunable Empty Pipe Parameters.
  • Page 8 Rosemount Sensors ........
  • Page 9: System Description

    The transmitter can be integrally or remotely mounted from the sensor. This manual is designed to assist in the installation and operation of the Rosemount 8712 Magnetic Flowmeter Transmitter and the Rosemount 8700 Series Magnetic Flowmeter Sensors. www.rosemount.com...
  • Page 10: Safety Messages

    Attempting to install and operate the Rosemount 8705, Rosemount 8707 High-Signal, Rosemount 8711, or Rosemount 8721 Magnetic Sensors with the Rosemount 8712 or Rosemount 8732 Magnetic Flowmeter Transmitter without reviewing the instructions contained in this manual could result in personal injury or equipment damage.
  • Page 11: Safety Messages

    Verify that the operating environment of the sensor and transmitter is consistent with the appropriate hazardous area approval. Do not connect a Rosemount 8712 to a non-Rosemount sensor that is located in an explosive atmosphere. www.rosemount.com...
  • Page 12: Transmitter Symbols

    Installation of this transmitter in an explosive environment must be in accordance with the appropriate local, national, and international standards, codes, and practices. Please review the approvals section of the 8712 reference manual for any restrictions associated with a safe installation.
  • Page 13 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure 2-1. Rosemount 8712 Dimensional Drawing WITH STANDARD COVER 4.31 9.01 (109) (229) 0.44 2.81 3.11 (11) (71) (79) 12.02 (305) 11.15 (283) 2.96 (75) WITH LOI COVER –14 NPT Conduit...
  • Page 14: Environmental Considerations

    Attach the 8712 to the mounting location using the mounting screws. Identify Options and The standard application of the 8712 includes a 4–20 mA output and control of the sensor coils. Other applications may require one or more of the...
  • Page 15 Connect 4–20 mA Loop External Power Source on page 2-9. Transmitter Security The security switch on the 8712 allows the user to lock out any configuration changes attempted on the transmitter. No changes to the configuration are allowed when the switch is in the ON position. The flow rate indication and totalizer functions remain active at all times.
  • Page 16: Conduit Ports And Connections

    AC or DC power wiring. • Device must be properly grounded or earthed according to local electric codes. • Rosemount combination cable model number 08712-0752-0001 (ft) or 08712-0752-0003 (m) is required to be used to meet EMC requirements.
  • Page 17: Electrical Considerations

    Transmitter Input Power The 8712 transmitter is designed to be powered by 90-250 V AC, 50–60 Hz or 12–42 V DC. The eight digit in the transmitter model number designates the appropriate power supply requirement.
  • Page 18 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table 2-1. Length of Annealed Copper (cu) Wires Types of Power Maximum Length of the Wire for Each Supply Wires Corresponding Power Supply Source Wire Annealed Cu 42 V DC 30 V DC 20 V DC 12.5 V DC...
  • Page 19: Installation Category

    3 Amp, Quick Acting Bussman AGC3 or Equivalent OPTIONS, If the application of the 8712 includes the use of options such as multidrop communications, auxiliary output control, or pulse output, certain CONSIDERATIONS, AND requirements may apply in addition to those previously listed. Be prepared to...
  • Page 20: Connect Pulse Output Power Source

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Internal The 4–20 mA analog power loop may be powered from the transmitter itself. Resistance in the loop must be 1,000 ohms or less. If a Handheld Communicator or control system will be used, it must be connected across a minimum of 250 ohms resistance in the loop.
  • Page 21: Connect Auxiliary Channel 1

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 The pulse output option requires an external power source. Complete the following steps to connect an external power supply. Ensure that the power source and connecting cable meet the requirements outlined previously.
  • Page 22: Connect Auxiliary Channel 2

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 When configured as an output, the following requirements apply: Supply Voltage: 5 to 28V DC Maximum Power: 2 watts Switch Closure: optically isolated solid state switch When using channel 1 as a digital output, the power source must be connected to the transmitter.
  • Page 23: Sensor Connections

    This section covers the steps required to physically install the transmitter including wiring and calibration. Rosemount Sensors To connect the transmitter to a non-Rosemount sensor, refer to the appropriate wiring diagram in Appendix D: Wiring Diagrams. The calibration procedure listed is not required for use with Rosemount sensors.
  • Page 24: Conduit Cables

    For remote mount installations, combination signal and coil drive cable should be limited to less than 300 ft. (100 m). Rosemount recommends using the combination signal and coil drive for N5, E5 approved sensors for optimum performance.
  • Page 25: Sensor To Remote Mount Transmitter Connections

    Connect coil drive and electrode cables as shown in Figure 2-13. Transmitter Connections Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary. Figure 2-13. Wiring Diagram Rosemount 8712 Transmitter Rosemount 8705/8707/8711/8721 sensors 2-15...
  • Page 26: Rev Ba Rosemount

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 2-16...
  • Page 27: Introduction

    “Universal Sensor Wiring Diagrams” on page E-1. The Rosemount 8712 features a full range of software functions for configuration of output from the transmitter. Software functions are accessed through the LOI, AMS, a Handheld Communicator, or a control system.
  • Page 28: Local Operator Interface

    Refer to Section 6 "Maintenance and Troubleshooting" for further information. LOCAL OPERATOR The optional Local Operator Interface (LOI) provides an operator communications center for the 8712. By using the LOI, the operator can INTERFACE access any transmitter function for changing configuration parameter settings, checking totalized values, or other functions.
  • Page 29: Basic Features

    Transmitter Parameter Keys The transmitter parameter keys provide direct access to the most common transmitter parameters and stepped access to the advanced functions of the 8712 through the AUX. FUNCTION key. Figure 3-1. Local Operator Interface Keypad DATA...
  • Page 30: Selecting Options

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Selecting Options To select pre-defined software options on the LOI, use the following procedure: Access the appropriate option. Use SHIFT or INCR. to toggle between the applicable choices. Press ENTER when the desired choice is displayed on the screen.
  • Page 31 Function Performed Keys Tube Cal No. Identifies the calibration number when using Rosemount sensors, or other manufacturers’ sensors calibrated at the Rosemount factory Tube Size Specifies the sensor size and identifies the corresponding maximum flow (0.1 - through 80-inch line sizes)
  • Page 32: Diagnostic Messages

    • Totalizer indicator (A flashing letter “T” on the LOI indicates to totalizer is activated) Review The 8712 includes a capability that enables you to review the configuration variable settings. Fast Keys 1, 5 The flowmeter configuration parameters set at the factory should be reviewed to ensure accuracy and compatibility with your particular application of the flowmeter.
  • Page 33: Pv - Primary Variable

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 PV - Primary Variable The Primary Variable shows the current measured flow rate. This value determines the analog output from the transmitter. Fast Keys 1, 1, 1 LOI Key FLOW RATE...
  • Page 34: Pulse Output

    The Pulse Output displays the current value of the pulse signal. Fast Keys 1, 1, 5 BASIC SETUP The basic configuration functions of the Rosemount 8712 must be set for all applications of the transmitter in a magnetic flowmeter system. If your Fast Keys 1, 3 application requires the advanced functionality features of the Rosemount 8712, see Section 4 "Operation"...
  • Page 35 (see “Flow Units” on page 3-8). If your application has special needs and the standard configurations do not apply, the Rosemount 8712 provides the flexibility to configure the transmitter in a custom-designed units format using the special units variable.
  • Page 36: Line Size

    The actual special units setting you define will not appear. Four characters are available to store the new units designation. The 8712 LOI will display the four character designation as configured.
  • Page 37: Pv Urv (Upper Range Value)

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 PV URV The upper range value (URV), or analog output range, is preset to 30 ft/s at the factory. The units that appear will be the same as those selected under the (Upper Range Value) units parameter.
  • Page 38: Calibration Number

    Rosemount flow lab, accuracy of the system may be compromised. If your sensor is not a Rosemount sensor and was not calibrated at the Rosemount factory, contact your Rosemount representative for assistance. If your sensor is imprinted with an eight-digit number or a k-factor, check in the sensor wiring compartment for the sixteen-digit calibration number.
  • Page 39: Introduction

    This section contains information for advanced configuration parameters and diagnostics. The software configuration settings for the Rosemount 8712 can be accessed through a HART-based communicator, Local Operator Interface (LOI), or through a control system. The software functions for the Field Communicator are described in detail in this section of the manual.
  • Page 40: Basic Diagnostics

    For more details on the electronics temperature diagnostic, see Appendix C: "Diagnostics". Basic Diagnostics The basic diagnostics menu contains all of the standard diagnostics and tests that are available in the 8712 transmitter. Field Comm. 1, 2, 2 Self Test Field Comm.
  • Page 41 1, 2, 2, 2, 3 Send the analog output into an alarm mA value. Actual mA value depends on the alarm configuration. • Rosemount Standard High Alarm – 22.6 mA • Rosemount Standard Low Alarm – 3.75 mA • Namur Compliant High Alarm – 22.6 mA •...
  • Page 42 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Empty Pipe Trigger Level Field Comm. 1, 2, 2, 4, 2 LOI Key AUX. FUNCTION Limits: 3 to 2000 Configure the threshold limit that the empty pipe value must exceed before the diagnostic alert activates.
  • Page 43 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 In Range – The HART Alert will activate when the measured flow rate is between the High Limit 1 and Low Limit 1 set points. Out of Range – The HART Alert will activate when the measured flow rate exceeds the High Limit 1 set point or falls below the Low Limit 1 set point.
  • Page 44 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 High Limit 2 Field Comm. 1, 2, 2, 7, 3 Set the flow rate value that corresponds to the high limit set point for the Flow Limit 2 alert. If a digital output is configured for Flow Limit 1, the digital output will activate when the conditions for mode 1 are met.
  • Page 45: Advanced Diagnostics

    Alert status. Advanced Diagnostics The advanced diagnostics menu contains information on all of the additional diagnostics and tests that are available in the 8712 transmitter if one of the Field Comm. 1, 2, 3 diagnostics suite packages was ordered.
  • Page 46 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 8714i Results Field Comm. 1, 2, 3, 1, 2 LOI Key XMTR INFO Review the results of the most recently performed 8714i Meter Verification test. Information in this section details the measurements taken and if the meter passed the verification test.
  • Page 47 Have the transmitter measure and store the sensor signature values. These values will then be used as the baseline for the meter verification test. Use this when connecting to older Rosemount or another manufacturers sensors, or installing the magnetic flowmeter system for the first time. For more details...
  • Page 48 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Recall Last Saved Values Field Comm. 1, 2, 3, 1, 3, 3 LOI Key AUX. FUNCTION Recalls the last saved “good” values for the sensor signature. Set Pass/Fail Criteria Field Comm.
  • Page 49 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Coil Signature Field Comm. 1, 2, 3, 1, 5, 2 View the measured value for the coil signature taken during the 8714i meter verification test. Electrode Resistance Field Comm. 1, 2, 3, 1, 5, 3 View the measured value for the electrode resistance taken during the 8714i meter verification test.
  • Page 50: Diagnostic Variable Values

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Diagnostic Variable From this menu, all of the diagnostic variable values can be reviewed. This information can be used to get more information about the transmitter, sensor, Values and process, or to get more detail about an alert that may have activated.
  • Page 51 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 8714i Test Result Field Comm. 1, 2, 3, 7, 3 Displays the results of the 8714i Meter Verification test as pass or fail. For more details on this parameter see Appendix C: "Diagnostics".
  • Page 52: Trims

    Digital trim is the function by which the factory calibrates the transmitter. This procedure is rarely needed by users. It is only necessary if you suspect the Rosemount 8712 is no longer accurate. A Rosemount 8714D Calibration Standard is required to complete a digital trim. Attempting a digital trim without a Rosemount 8714D Calibration Standard may result in an inaccurate transmitter or an error message.
  • Page 53 “DIGITAL TRIM FAILURE” message may appear. If this message occurs, no values were changed in the transmitter. Simply power down the Rosemount 8712 to clear the message. To simulate a nominal sensor with the Rosemount 8714D Calibration...
  • Page 54: Status

    LOI Key XMTR INFO ADVANCED In addition to the basic configuration options and the diagnostic information and controls, the 8712 has many advanced functions that can also be CONFIGURATION configured as required by the application. DETAILED SETUP The detailed setup function provides access to other parameters within the transmitter that can be configured such as coil drive frequency, output Field Comm.
  • Page 55: Configure Outputs

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Density Value Field Comm. 1, 4, 1, 2 LOI Key AUX. FUNCTION The density value is used to convert from a volumetric flow rate to a mass flow rate using the following equation:...
  • Page 56 Field Comm. 1, 4, 2, 1, 4 The analog output alarm type displays the alarm mode the 8712 is currently set for. This value is set by a switch on the electronics board. There are two available options for this setting: •...
  • Page 57 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Loop Test Field Comm. 1, 4, 2, 1, 5 LOI Key AUX. FUNCTION The loop test allows you to drive the transmitter output to a desired electrical current output on terminals 1 and 2. This capability allows you to check the entire current loop prior to start-up.
  • Page 58 10,000 Hz. With the 110 percent overrange capability, the absolute limit is 11,000 Hz. For example, if you want the Rosemount 8712 to pulse every time 0.01 gallons pass through the sensor, and the flow rate is 10,000 gal/min, you will exceed the 10,000 Hz full-scale limit: ...
  • Page 59 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 The best choice for this parameter depends upon the required resolution, the number of digits in the totalizer, the extent of range required, and the maximum counter external frequency. NOTE For totalizing on the LOI, ten digits are available.
  • Page 60 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Example The maximum flow rate is 10,000 gpm. Set the pulse output scaling such that the transmitter outputs 10,000 Hz at 10,000 gpm. Flow Rate (gpm) Pulse Scaling ---------------------------------------------------------- - (60 s/min)(Frequency)
  • Page 61 AUX. FUNCTION This menu is used to configure the optional digital input and digital output parameters of the 8712 transmitter. Note that this configuration option is only active if the auxiliary output suite (option code AX) was ordered or licensed in the field.
  • Page 62 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 DIO 1 Control Field Comm. 1, 4, 2, 3, 1, 2 Displays the configuration for Channel 1 as either a discrete Input or Output. Digital Input 1 Field Comm. 1, 4, 2, 3, 1, 3 Displays what digital input Channel 1 will be set to when the Control for Channel 1 is set to Input.
  • Page 63 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Control 1 Field Comm. 1, 4, 2, 3, 3, 1 Turns the Flow Limit 1 HART Alert ON or OFF. ON – The transmitter will generate a HART alert when the defined conditions are met.
  • Page 64 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 ON – The transmitter will generate a HART alert when the defined conditions are met. If a digital output is configured for Flow Limit 1, the digital output will activate when the conditions for mode 1 are met.
  • Page 65 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Total Mode Field Comm. 1, 4, 2, 3, 5, 2 Mode that determines when the Total Limit HART Alert will activate. > High Limit – The HART Alert will activate when the measured net total exceeds the Total High Limit set point.
  • Page 66 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Totalizer Units Field Comm. 1, 4, 2, 5, 1 LOI Key AUX. FUNCTION Totalizer units allow for the configuration of the units that the totalized value will be displayed as. These units are independent of the flow units.
  • Page 67 The alarm level allows you to drive the transmitter to preset values if an alarm occurs. There are two options: • Rosemount Alarm and Saturation Values • NAMUR-Complaint Alarm and Saturation Levels Table 4-3. Rosemount (Standard) Alarm and Saturation Values Level 4-20 mA Saturation 4-20 mA Alarm 3.75 mA 3.9 mA 22.6 mA...
  • Page 68 Rosemount 8712 in the loop. NOTE The Rosemount 8712 is set to poll address zero at the factory, allowing it to operate in the standard point-to-point manner with a 4–20 mA output signal. To activate multidrop communication, the transmitter poll address must be changed to a number between 1 and 15.
  • Page 69: Loi Configuration

    Forward Total and Reverse Total • Net Total and Gross Total Signal Processing The 8712 contains several advanced functions that can be used to stabilize erratic outputs caused by process noise. The signal processing menu Field Comm. 1, 4, 4 LOI Key AUX.
  • Page 70 Field Comm. 1, 4, 4, 2, 1 When ON is selected, the Rosemount 8712 output is derived using a running average of the individual flow inputs. Signal processing is a software algorithm that examines the quality of the electrode signal against user-specified tolerances.
  • Page 71 10 seconds. In some cases this may be unacceptable. By setting the time limit, you can force the 8712 to clear the value of the running average and re-establish the output and average at the new flow rate once the time limit has elapsed.
  • Page 72: Universal Auto Trim

    00809-0100-4664, Rev BA Rosemount 8712 January 2010 Universal Auto Trim The universal auto trim function enables the Rosemount 8712 to calibrate sensors that were not calibrated at the Rosemount factory. The function is Field Comm. 1, 4, 5 LOI Key AUX.
  • Page 73 Revisions numbers are fixed informational variables that provide the revision number for different elements of your Field Communicator and Rosemount 8712. These revision numbers may be required when calling the factory for support. Revision numbers can only be changed at the factory and are...
  • Page 74 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Software Revision Number Field Comm. 1, 4, 6, 10, 3 This function displays the software revision number of the transmitter. This is one piece of information required to generate a license code to enable diagnostics in the field.
  • Page 75 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Flange Material Field Comm. 1, 4, 6, 11, 2 LOI Key XMTR INFO Flange material enables you to select the flange material for your magnetic transmitter system. This variable only needs to be changed if you have changed your sensor.
  • Page 76 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Liner Material Field Comm. 1, 4, 6, 11, 5 LOI Key XMTR INFO Liner material enables you to select the liner material for the attached sensor. This variable only needs to be changed if you have replaced your sensor.
  • Page 77 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure 4-2. Field Communicator Menu Tree for the Rosemount 8712 4-39...
  • Page 78 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure 4-3. HART Fast Key Function HART Fast Keys Process Variables (PV) Primary Variable Value 1,1,1 Primary Variable % 1,1,2 PV Loop Current 1,1,3 Totalizer Set-Up 1,1,4 Totalizer Units 1,1,4,1 Gross Total...
  • Page 79 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Function HART Fast Keys Sensor Cal Test Result 1,2,3,1,2,9 Coil Circuit Test Result 1,2,3,1,2,x Electrode Circuit Test Result 1,2,3,1,2,x Sensor Signature 1,2,3,1,3 Signature Values 1,2,3,1,3,1 Coil Resistance 1,2,3,1,3,1,1 Coil Signature 1,2,3,1,3,1,2...
  • Page 80 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Function HART Fast Keys Base Volume Unit 1,3,2,2,2 Conversion Number 1,3,2,2,3 Base Time Unit 1,3,2,2,4 Flow Rate Unit 1,3,2,2,5 Line Size 1,3,3 PV URV 1,3,4 PV LRV 1,3,5 Calibration Number 1,3,6...
  • Page 81 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Function HART Fast Keys Diagnostic Status Alert 1,4,2,3,6 Reverse Flow 1,4,2,4 Totalizer Setup 1,4,2,5 Totalizer Units 1,4,2,5,1 Gross Total 1,4,2,5,2 Net Total 1,4,2,5,5 Reverse Total 1,4,2,5,4 Start Totalizer 1,4,2,5,5 Stop Totalizer...
  • Page 82 Function Performed Keys Tube Cal No. Identifies the calibration number when using Rosemount sensors, or other manufacturers’ sensors calibrated at the Rosemount factory Tube Size Specifies the sensor size and identifies the corresponding maximum flow (0.1 - through 80-inch line sizes)
  • Page 83 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Auxiliary Functions Function Options Run 8714i Runs the meter verification diagnostic Operating Mode Normal or Filter Coil Pulse Mode 5 or 37 Hz Flow rate Display Flow–% Span, Flow–Totalize, %Span–Totalize Totalizer Display Forward–Reverse or Net–Gross...
  • Page 84 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 4-46...
  • Page 85: Safety Messages

    Verify that the operating environment of the sensor and transmitter is consistent with the appropriate hazardous area approval. Do not connect a Rosemount 8712 to a non-Rosemount sensor that is located in an explosive atmosphere. www.rosemount.com...
  • Page 86 Installation of this transmitter in an explosive environment must be in accordance with the appropriate local, national, and international standards, codes, and practices. Please review the approvals section of the 8712 reference manual for any restrictions associated with a safe installation.
  • Page 87: Sensor Handling

    Figure 5-1 shows sensors correctly supported for handling and installation. Notice the plywood end pieces are still in place to protect the sensor liner during transportation. Figure 5-1. Rosemount 8705 Sensor Support for Handling 6-Inch and Larger ½- through 4-Inch...
  • Page 88: Sensor Mounting

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 SENSOR MOUNTING Physical mounting of a sensor is similar to installing a typical section of pipe. Conventional tools, equipment, and accessories (bolts, gaskets, and grounding hardware) are required. Upstream/Downstream To ensure specification accuracy over widely varying process conditions,...
  • Page 89 Orientation FLOW The electrodes in the Rosemount 8711 are properly oriented when the top of the sensor is either vertical or horizontal, as shown in Figure 5-6. Avoid any mounting orientation that positions the top of the sensor at 45 degrees from...
  • Page 90: Flow Direction

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure 5-6. Rosemount 8711 Mounting Position 45° Electrode Plane 45° Electrode Plane Flow Direction The sensor should be mounted so that the FORWARD end of the flow arrow, shown on the sensor identification tag, points in the direction of flow through the tube (see Figure 5-7).
  • Page 91: Installation (Flanged Sensor)

    The following section should be used as a guide in the installation of the flange-type Rosemount 8705 and Rosemount 8707 High-Signal Sensors. (FLANGED SENSOR) Refer to page 5-10 for installation of the wafer-type Rosemount 8711 Sensor. Gaskets The sensor requires a gasket at each of its connections to adjacent devices or piping.
  • Page 92 All sensors require a second torquing 24 hours after initial flange bolt tightening. Table 5-1. Flange Bolt Torque Specifications for Rosemount 8705 and 8707 High-Signal Sensors PTFE/ETFE liner Polyurethane liner Class 150...
  • Page 93 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table 5-2. Flange Bolt Torque and Bolt Load Specifications for Rosemount 8705 PTFE/ETFE liner PN10 PN 16 PN 25 PN 40 Size Code Line Size (Newton-meter) (Newton) (Newton-meter) (Newton) (Newton-meter) (Newton) (Newton-meter) (Newton)
  • Page 94: Installation (Wafer Sensor)

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table 5-3. Flange Bolt Torque and Bolt Load Specifications for Rosemount 8705 Polyurethane Liner PN 10 PN 16 PN 25 PN 40 Size Code Line Size (Newton-meter) (Newton) (Newton-meter) (Newton) (Newton-meter) (Newton) (Newton-meter) (Newton)
  • Page 95: Flange Bolts

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table 5-4. Stud Specifications Nominal Sensor Size Stud Specifications 0.15 – 1 inch (4 – 25 mm) 316 SST ASTM A193, Grade B8M Class 1 threaded mounted studs – 8 inch (40 – 200 mm) CS, ASTM A193, Grade B7, threaded mounting studs Place the sensor between the flanges.
  • Page 96: Installation (Sanitary Sensor)

    The sensor requires a gasket at each of its connections to adjacent devices or piping. The gasket material selected must be compatible with the process fluid and operating conditions. Gaskets are supplied with all Rosemount 8721 Sanitary sensors except when the process connection is an IDF sanitary screw type.
  • Page 97 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 NOTE Consult factory for installations requiring cathodic protection or situations where there are high currents or high potential in the process. The sensor case should always be earth grounded in accordance with national and local electrical codes.
  • Page 98 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure 5-14. Grounding with Grounding Rings or Lining Protectors Grounding Rings or Lining Protectors Figure 5-15. Grounding with Grounding Rings or Lining Protectors Grounding Rings or Lining Protectors 5-14...
  • Page 99: Process Leak Protection (Optional)

    Figure 5-16. Grounding with Grounding Electrodes PROCESS LEAK The Rosemount 8705 and 8707 High-Signal Sensor housing is fabricated from carbon steel to perform two separate functions. First, it provides PROTECTION shielding for the sensor magnetics so that external disturbances cannot (OPTIONAL) interfere with the magnetic field and thus affect the flow measurement.
  • Page 100: Relief Valves

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure 5-17. Standard Housing Configuration — Sealed Welded Housing (Option Code W0) –14 NPT Conduit Connection (no relief valve) Relief Valves The first optional configuration, identified by the W1 in the model number option code, uses a completely welded coil housing.
  • Page 101: Process Leak Containment

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Process Leak The second optional configuration, identified as option code W3 in the model number, divides the coil housing into three compartments: one for each Containment electrode and one for the coils. Should a damaged liner or electrode fault allow process fluid to migrate behind the electrode seals, the fluid is contained in the electrode compartment.
  • Page 102 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 5-18...
  • Page 103: Safety Information

    FM or CSA approval. Do not connect a Rosemount 8712 to a non-Rosemount sensor that is located in an explosive atmosphere. Mishandling products exposed to a hazardous substance may result in death or serious injury.
  • Page 104: Installation Check And Guide

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 INSTALLATION CHECK Use this guide to check new installations of Rosemount magnetic flowmeter systems that appear to malfunction. AND GUIDE Before You Begin Transmitter Apply power to your system before making the following transmitter checks.
  • Page 105: Diagnostic Messages

    Problems in the magnetic flowmeter system are usually indicated by incorrect output readings from the system, error messages, or failed tests. Consider all MESSAGES sources in identifying a problem in your system. Table 6-1. Rosemount 8712 Basic Diagnostic Messages Message Potential Cause Corrective Action “Empty Pipe”...
  • Page 106 Calibrator is not set to 30 FPS Change calibrator setting to 30 FPS Bad calibrator Replace calibrator Table 6-2. Rosemount 8712 Advanced Diagnostic Messages (Suite 1 - Option Code DA1) Message Potential Cause Corrective Action Grounding/Wiring Fault Improper installation of wiring See “Sensor to Remote Mount Transmitter Connections”...
  • Page 107 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table 6-4. Basic Troubleshooting–Rosemount 8712 Symptom Potential Cause Corrective Action Output at 0 mA No power to transmitter Check power source and connections to the transmitter Blown fuse Check the fuse and replace with an appropriately rated fuse, if necessary...
  • Page 108: Transmitter Troubleshooting

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 TRANSMITTER TROUBLESHOOTING Table 6-5. Advanced Troubleshooting–Rosemount 8712 Symptom Potential Cause Corrective Action Does not appear to be within Transmitter, control system, or other Check all configuration variables for the transmitter, sensor,...
  • Page 109 Sensor calibration number Units Line size Electrode coating Use bulletnose electrodes in the Rosemount 8705 Sensor. Downsize the sensor to increase the flow rate above 3 ft/s. Periodically clean the sensor Air in line Move the sensor to another location in the process line to...
  • Page 110: Quick Troubleshooting

    To interpret the results, the hazardous location certification for the sensor must be known. Applicable codes for the Rosemount 8705 are N0, N5, and KD. Applicable codes for the Rosemount 8707 are N0 and N5. Applicable codes for the Rosemount 8711 are N0, N5, E5, and CD.
  • Page 111 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table 6-6. Sensor Test Sensor Required Measuring at Test Location Equipment Connections Expected Value Potential Cause Corrective Action   A. Sensor Installed or Multimeter 1 and 2 = R 2...
  • Page 112: Step 4: Uninstalled Sensor Tests

    Sensor Tests be known. Applicable codes for the Rosemount 8705 are N0, N5, and KD. Applicable codes for the Rosemount 8707 are N0 and N5. Applicable codes for the Rosemount 8711 are N0, N5, E5, and CD.
  • Page 113 Measure both electrodes. One electrode should result in an open reading, while  the other electrode should be less than 275 Table 6-8. Uninstalled Rosemount 8711 Wafer Sensor Tests Hazardous Location Certification Measuring at Connections N5, E5, CD 18 and Electrode ...
  • Page 114 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 6-12...
  • Page 115: Functional Specifications

    Physical Specifications ......page A-8 Rosemount 8712E Ordering Information ... . . page A-9...
  • Page 116 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 DC Load Limitations (Analog Output) Maximum loop resistance is determined by the voltage level of the external power supply, as described by: FIGURE 1. DC Load Limitations 1250 1000 Operating Region 10.8...
  • Page 117 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Humidity Limits 0–100% RH to 120 °F (49 °C), decreases linearly to 10% RH at 130 °F (54 °C) Enclosure Rating Type 4X, IP66 Output Signals Analog Output Adjustment 4–20 mA, switch-selectable as internally or externally powered 5 to 24 V DC;...
  • Page 118 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Flow Limits (2): Activates switch closure output when the transmitter measures a flow rate that meets the conditions established for this alert. There are two independent flow limit alerts that can be configured as discrete outputs.
  • Page 119 Damping Adjustable between 0.0 and 256 seconds Sensor Compensation Rosemount sensors are flow-calibrated and assigned a calibration factor at the factory. The calibration factor is entered into the transmitter, enabling interchangeability of sensors without calculations or a compromise in accuracy.
  • Page 120: Performance Specifications

    Includes the combined effects of linearity, hysteresis, repeatability, and calibration uncertainty. Rosemount 8712E with 8705/8707 Sensor: Standard system accuracy is ±0.25% of rate ±1.0 mm/sec from 0.04 to 6 ft/s (0.01 to 2 m/s); above 6 ft/s (2 m/s), the system has an accuracy of ±0.25% of rate ±1.5 mm/sec.
  • Page 121 0.04 and 3.0 ft/s (0.01 and 1 m/s), the system has an accuracy of ±0.015 ft/s (0.005 m/s). Rosemount 8712E with Other Manufacturers’ Sensors: When calibrated in the Rosemount Flow Facility, system accuracies as good as 0.5% of rate can be attained. There is no accuracy specification for other manufacturers’ sensors calibrated in the process line.
  • Page 122: Physical Specifications

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 PHYSICAL Materials of Construction SPECIFICATIONS Housing Low-copper aluminum, Type 4X and IEC 60529 IP66 Paint Polyurethane Cover Gasket Rubber Electrical Connections Four –14 NPT connections provided on the base of the transmitter. Screw terminals provided for all of the connections.
  • Page 123: Rosemount 8712E Ordering Information

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 ROSEMOUNT 8712E ORDERING INFORMATION Model Product Description 8712E Remote Magnetic Flowmeter Transmitter Code Transmitter Style Standard Code Transmitter Mount Remote Mount for 2 in. pipe or panel (includes CS mounting bolts and 316L SST bracket)
  • Page 124 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 QIG Language Danish Dutch French German Finnish Italian Norwegian Portuguese Spanish Russian Swedish Typical Model Number: 8712E S R 1 A 1 N0 DA1 DA2 M4 (1) Adapter are used for this conduit entry type...
  • Page 125: Product Certifications

    PRODUCT CERTIFICATIONS APPROVED Rosemount Inc. — Eden Prairie, Minnesota, USA MANUFACTURING Fisher-Rosemount Technologias de Flujo, S.A. de C.V. — Chihuahua Mexico LOCATIONS Emerson Process Management Flow — Ede, The Netherlands Asia Flow Technology Center — Nanjing, China European Directive The EC declaration of conformity can be found in document number 00825-0100-4664.
  • Page 126: Sensor Approval Information

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Canadian Standards Association (CSA) Non-incendive for Class I, Division 2, Groups A, B, C, and D non-flammable fluids (T4 at 40 °C), and Dust-ignition proof Class II/III, Division 1, Groups E, F, and G (T4 at 40 °C) Hazardous locations;...
  • Page 127 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 North American Certifications Factory Mutual (FM) Non-incendive for Class I, Division 2, Groups A, B, C, and D non-flammable fluids (8705/8711 T5 at 60 °C; 8707 T3C at 60 °C), and Dust-ignition proof Class II/III, Division 1, Groups E, F, and G (8705/8711 T6 at 60 °C;...
  • Page 128 A fuse with a rating of maximum 0,7 A according to IEC 60127-1 shall be included in the coil excitation circuit if the sensors are used with other flow transmitters. Table B-1. Electrical Data Rosemount 8705 and 8711 Sensors Coil excitation 40 V, 0,5 A, 20 W maximum...
  • Page 129 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table B-3. Relation between the maximum ambient temperature, the maximum process temperature, and the temperature class Maximum Maximum process temperature °F (°C) per temperature class Ambient Temperature 0.5 in. sensor size 149°F (65°C)
  • Page 130 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table B-3. Relation between the maximum ambient temperature, the maximum process temperature, and the temperature class Maximum Maximum process temperature °F (°C) per temperature class Ambient Temperature 2.0 in. sensor size 149°F (65°C)
  • Page 131 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure B-1. CSA Installation...
  • Page 132 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure B-2. Factory Mutual Hazardous Locations...
  • Page 133: Diagnostic Availability

    Rosemount magmeter diagnostics can be accessed through the Local Operator Interface (LOI), the 375 Handheld Communicator, and AMS Device Manager. Access Diagnostics through the LOI for quicker installation, maintenance, and meter verification Rosemount magmeter diagnostics are available through the LOI to make maintenance of every magmeter easier. www.rosemount.com...
  • Page 134: Licensing And Enabling

    ENABLING diagnostics can be licensed in the field through the use of a license key. To obtain a license key, contact your local Rosemount Representative. Each transmitter has a unique license key specific to the diagnostic option code. See the detailed procedures below for entering the license key and enabling the advanced diagnostics.
  • Page 135: Tunable Empty Pipe Parameters

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Tunable Empty Pipe The Tunable Empty Pipe diagnostic has one read-only parameter, and two parameters that can be custom configured to optimize the diagnostic Parameters performance. Empty Pipe Value HART Fast Keys...
  • Page 136: Troubleshooting Empty Pipe

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Example Set the counts to 10 Troubleshooting Empty The following actions can be taken if Empty Pipe detection is unexpected. Pipe Verify the sensor is full. Verify that the sensor has not been installed with a measurement electrode at the top of the pipe.
  • Page 137: Troubleshooting Ground/Wiring Fault

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Troubleshooting The transmitter detected high levels of 50/60 Hz noise caused by improper wiring or poor process grounding. Ground/Wiring Fault Verify that the transmitter is earth grounded. Connect ground rings, grounding electrode, lining protector, or grounding straps.
  • Page 138: High Process Noise Parameters

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 High Process Noise The High Process Noise diagnostic has two read-only parameters. It does not have any configurable parameters. This diagnostic requires that flow be Parameters present in the pipe and the velocity be > 1 ft/s.
  • Page 139: High Process Noise Functionality

    Rosemount High-Signal 8707 sensor be used. These sensors can be calibrated to run at lower coil drive current supplied by the standard Rosemount transmitters, but can also be upgraded by changing to the 8712H High-Signal transmitter.
  • Page 140: 8714I Meter Verification

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 The transmitter continuously monitors signal amplitudes over a wide range of frequencies. For the high process noise diagnostic, the transmitter specifically looks at the signal amplitude at frequencies of 2.5 Hz, 7.5 Hz, 32.5 Hz, and 42.5 Hz.
  • Page 141: 8714I Meter Verification Test Parameters

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 8714i Meter Verification The 8714i has a multitude of parameters that set the test criteria, test conditions, and scope of the calibration verification test. Test Parameters Test Conditions for the 8714i Meter Verification There are three possible test conditions that the 8714i Meter Verification test can be initiated under.
  • Page 142: 8714I Meter Verification Test Results Parameters

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Empty Pipe Set the test criteria for the Empty Pipe condition. The factory default for this value is set to three percent with limits configurable between one and ten percent. HART Fast Keys...
  • Page 143 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Viewing the 8714i Meter Verification Results Depending on the method used to view the results, they will be displayed in either a menu structure, as a method, or in the report format. When using the HART Field Communicator, each individual component can be viewed as a menu item.
  • Page 144 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Actual Velocity Displays the velocity measured by the transmitter during the transmitter calibration verification process. HART Fast Keys 1,2,3,3,2,5 LOI Key XMTR INFO Context Menu, Device Diagnostics, 8714i Report Velocity Deviation Displays the deviation in the actual velocity compared to the simulated velocity in terms of a percentage.
  • Page 145: Optimizing The 8714I Meter Verification

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Optimizing the 8714i The 8714i Meter Verification diagnostic can be optimized by setting the test criteria to the desired levels necessary to meet the compliance requirements Meter Verification of the application. The following examples below will provide some guidance on how to set these levels.
  • Page 146: Troubleshooting The 8714I Meter Verification Test

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Troubleshooting the In the event that the 8714i Meter Verification test fails, the following steps can be used to determine the appropriate course of action. Begin by reviewing the 8714i Meter Verification 8714i results to determine the specific test that failed.
  • Page 147 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Coil Signature The Coil Signature is a measurement of the magnetic field strength. This value is used as a baseline to determine if a sensor calibration shift has occurred when the 8714i Meter Verification diagnostic is initiated.
  • Page 148: Rosemount Magnetic Flowmeter Calibration Verification Report

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 ROSEMOUNT MAGNETIC FLOWMETER CALIBRATION VERIFICATION REPORT Calibration Verification Report Parameters Calibration Conditions: □ Internal □ External User Name: _____________________________________________ Test Conditions: □ Flowing □ No Flow, Full Pipe □ Empty Pipe...
  • Page 149: Safety Messages

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Appendix D Digital Signal Processing Safety Messages ....... . . page D-1 Procedures .
  • Page 150: Procedures

    00809-0100-4664, Rev BA Rosemount 8712 January 2010 PROCEDURES If the output of your Rosemount 8712 is unstable, first check the wiring and grounding associated with the magnetic flowmeter system. Ensure that the following conditions are met: • Ground straps are attached to the adjacent flange or ground ring? •...
  • Page 151 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 This software technique, known as signal processing, “qualifies” individual flow signals based on historic flow information and three user-definable parameters, plus an on/off control. These parameters are: Number of samples: The number of samples function sets the amount of time that inputs are collected and used to calculate the average value.
  • Page 152 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure D-1. Signal Processing Flow Rate Limit Time Limit 12 Samples = 1 Second Time Input flow signal from sensor. Average flow signals and transmitter output, determined by the “number of samples” parameter.
  • Page 153 January 2010 When Should Signal Processing Be Used? The Rosemount 8712 offers three separate functions that can be used in series for improving a noisy output. The first step is to toggle the coil drive to the 37 Hz mode and initialize with an auto zero. If the output is still noisy at this stage, signal processing should be actuated and, if necessary, tuned to match the specific application.
  • Page 154 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010...
  • Page 155 Diagrams Rosemount Sensors ......page E-3 ABB Sensors ........page E-7 Brooks Sensors .
  • Page 156 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Rosemount Transmitter Sensor Manufacturer Page Number Rosemount Rosemount 8712 Rosemount 8705, 8707, 8711, 8721 page E-3 Rosemount 8712 Rosemount 8701 page E-4 Brooks Rosemount 8712 Model 5000 page E-9 Rosemount 8712...
  • Page 157: Rosemount Sensors

    ROSEMOUNT SENSORS Rosemount Connect coil drive and electrode cables as shown in Figure E-1. 8705/8707/8711/8721 Sensors to Rosemount 8712 Transmitter Figure E-1. Wiring Diagram to a Rosemount 8712 Transmitter ROSEMOUNT 8712 TRANSMITTER 8705/8707/8711/8721 SENSORS Table E-1. Rosemount 8705/8707/8711/8721 Sensor Rosemount 8712 Transmitters...
  • Page 158 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Rosemount 8701 Sensor Connect coil drive and electrode cables as shown in Figure E-2 on page E-4. to Rosemount 8712 Transmitter Figure E-2. Wiring Diagram for Rosemount 8701 Sensor and ROSEMOUNT 8701 ROSEMOUNT 8712...
  • Page 159 Transmitter Figure E-4. Wiring Diagram for Rosemount 8711 Sensor and ROSEMOUNT 8711 Rosemount 8712 Transmitter SENSOR ROSEMOUNT 8712 TRANSMITTER Table E-2. Rosemount 8711 Sensor Wiring Connections Rosemount 8712 Rosemount 8711 Sensors Coils + Coils – Shield Electrode + Electrode –...
  • Page 160 Verify that the sensor coil is configured for series connection. Other manufacturers sensors may be wired in either a series or parallel circuit. All Rosemount magnetic sensors are wired in a series circuit. (Other manufacturers AC sensors (AC coils) wired for 220V operation are typically wired in parallel and must be rewired in series.)
  • Page 161: Abb Sensors

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 ABB SENSORS Connect coil drive and electrode cables as shown in Figure E-5. ABB Magmaster MFE and MFF Sensors (Old Version) to Rosemount 8712 Transmitter Figure E-5. Wiring Diagram for ABB Magmaster MFE and MFF...
  • Page 162 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 ABB Magmaster MFE Connect coil drive and electrode cables as shown in Figure E-6. and MFF Sensors (New Version) to Rosemount 8712 Transmitter Figure E-6. Wiring Diagram for ABB Magmaster MFE and MFF...
  • Page 163: Brooks Sensors

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 BROOKS SENSORS Connect coil drive and electrode cables as shown in Figure E-7. Model 5000 Sensor to Rosemount 8712 Transmitter Figure E-7. Wiring Diagram for Brooks Sensor Model 5000 and BROOKS MODEL...
  • Page 164 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Model 7400 Sensor to Connect coil drive and electrode cables as shown in Figure E-8. Rosemount 8712 Transmitter Figure E-8. Wiring Diagram for Brooks Sensor Model 7400 and BROOKS MODEL 7400...
  • Page 165: Endress And Hauser Sensors

    Endress and Hauser Promag 10/50/53/55 H/P/WS Sensors Figure E-9. Wiring Diagram for Endress and Hauser Sensors ENDRESS AND HAUSER and Rosemount 8712 SENSORS ROSEMOUNT 8712 TRANSMITTER Table E-7. Endress and Hauser Sensor Wiring Connections Rosemount 8712 Endress and Hauser Sensors This is a pulsed DC magnetic flowmeter.
  • Page 166 Connect coil drive and electrode cables as shown in Figure E-10. Promag 30/33/39 D/H/F Sensors (FS Versions) Figure E-10. Wiring Diagram for Endress and Hauser Sensors ENDRESS AND HAUSER and Rosemount 8712 SENSORS ROSEMOUNT 8712 TRANSMITTER Table E-8. Endress and Hauser Promag 30/33/39 D/H/F Sensors (FS Versions) Wiring Connections...
  • Page 167 Connect coil drive and electrode cables as shown in Figure E-11. Promag 30/33/39 A Sensors (FS Versions) Figure E-11. Wiring Diagram for Endress and Hauser Sensors ENDRESS AND HAUSER and Rosemount 8712 SENSORS ROSEMOUNT 8712 TRANSMITTER Table E-9. Endress and Hauser Promag 30/33/39 A Sensor (FS Versions) Wiring Connections...
  • Page 168 Connect coil drive and electrode cables as shown in Figure E-12. Generic Sensor to Rosemount 8712 Transmitter Figure E-12. Wiring Diagram for Endress and Hauser Sensors and Rosemount 8712 ENDRESS AND HAUSER GENERIC SENSORS ROSEMOUNT 8712 TRANSMITTER 42 42 41 41 Table E-10.
  • Page 169: Fischer And Porter Sensors

    Rosemount 8712 January 2010 FISCHER AND PORTER Connect coil drive and electrode cables as shown in Figure E-13. SENSORS Model 10D1418 Sensor to Rosemount 8712 Transmitter Figure E-13. Wiring Diagram for Fischer and Porter Sensor ROSEMOUNT Model 10D1418 and Rosemount...
  • Page 170 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Model 10D1419 Sensor Connect coil drive and electrode cables as shown in Figure E-14. to Rosemount 8712 Transmitter Figure E-14. Wiring Diagram for Fischer and Porter Sensor ROSEMOUNT Model 10D1419 and Rosemount...
  • Page 171 Connect coil drive and electrode cables as shown in Figure E-15. (Remote) to Rosemount 8712 Transmitter Figure E-15. Wiring Diagram for Fischer and Porter Sensor MODEL 10D1430 ROSEMOUNT 8712 TRANSMITTER Model 10D1430 (Remote) and SENSOR Rosemount 8712 Electrode Connections Coil Connections Table E-13. Fischer and Porter Model 10D1430 (Remote) Sensor...
  • Page 172 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Model 10D1430 Sensor Connect coil drive and electrode cables as shown in Figure E-16. (Integral) to Rosemount 8712 Transmitter Figure E-16. Wiring Diagram for Fischer and Porter Sensor MODEL 10D1430 ROSEMOUNT...
  • Page 173 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Model 10D1435 Sensor Connect coil drive and electrode cables as shown in Figure E-17. (Integral) to Rosemount 8712 Transmitter Figure E-17. Wiring Diagram for Fischer and Porter Sensor Model 10D1430 (Integral) and...
  • Page 174 Rosemount 8712 January 2010 Model 10D1465 and Connect coil drive and electrode cables as shown in Figure E-18. Model 10D1475 Sensors (Integral) to 8712 Transmitter Figure E-18. Wiring Diagram for Fischer and Porter Sensor MODEL 10D1465 AND MODEL ROSEMOUNT 8712...
  • Page 175: Fischer And Porter Sensors

    Fischer and Porter Connect coil drive and electrode cables as shown in Figure E-19. Generic Sensor to Rosemount 8712 Transmitter Figure E-19. Generic Wiring Diagram for Fischer and Porter Sensors and Rosemount 8712 ROSEMOUNT 8712 FISCHER AND PORTER TRANSMITTER SENSORS Electrodes Chassis...
  • Page 176: Foxboro Sensors

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 FOXBORO SENSORS Connect coil drive and electrode cables as shown in Figure E-20. Series 1800 Sensor to Rosemount 8712 Transmitter Figure E-20. Wiring Diagram for Foxboro Series 1800 and FOXBORO SERIES...
  • Page 177 Connect coil drive and electrode cables as shown in Figure E-21. Sensor to Rosemount 8712 Transmitter Figure E-21. Wiring Diagram for Foxboro Series 1800 (Version 2) ROSEMOUNT 8712 FOXBORO SERIES 1800 and Rosemount 8712 TRANSMITTER SENSOR (VERSION 2) White Black Shield Electrode Connections Coil Connections Table E-19.
  • Page 178 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Series 2800/8300 Sensor Connect coil drive and electrode cables as shown in Figure E-22. to 8712 Transmitter Figure E-22. Wiring Diagram for Foxboro Series 2800 and FOXBORO SERIES Rosemount 8712 2800/8300 SENSOR...
  • Page 179 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Series 8000A/9300A Connect coil drive and electrode cables as shown in Figure E-23. Sensor to 8712 Transmitter Figure E-23. Wiring Diagram for Foxboro Series 2800 and FOXBORO SERIES Rosemount 8712 ROSEMOUNT 8712...
  • Page 180 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Series 9100A/9200A Connect coil drive and electrode cables as shown in Figure E-24. Sensors to 8712 Transmitter Figure E-24. Wiring Diagram for Foxboro Series 9100A/9200A ROSEMOUNT 8712 and Rosemount 8712 TRANSMITTER...
  • Page 181 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Foxboro Generic Sensor Connect coil drive and electrode cables as shown in Figure E-25. to 8712 Transmitter Figure E-25. Generic Wiring Diagram for Foxboro Sensors FOXBORO and Rosemount 8712 SENSOR ROSEMOUNT 8712...
  • Page 182: Kent Sensors

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 KENT SENSORS Connect coil drive and electrode cables as shown in Figure E-26. Veriflux VTC Sensor to 8712 Transmitter Figure E-26. Wiring Diagram for Kent Veriflux VTC Sensor and KENT VERIFLUX VTC...
  • Page 183 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Kent Generic Sensor to Connect coil drive and electrode cables as shown in Figure E-27. Rosemount 8712 Transmitter Figure E-27. Generic Wiring Diagram for Kent Sensors and ROSEMOUNT 8712 Rosemount 8712...
  • Page 184: Krohne Sensors

    00809-0100-4664, Rev BA Rosemount 8712 January 2010 KROHNE SENSORS Connect coil drive and electrode cables as shown in Figure E-28. Krohne Autoflux Sensor to Rosemount 8712 Transmitter Figure E-28. Wiring Diagram for Krohne Autoflux Sensors and ROSEMOUNT 8712 Rosemount 8712...
  • Page 185 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Krohne Optiflux Sensor Connect coil drive and electrode cables as shown in Figure E-29. to Rosemount 8712 Transmitter Figure E-29. Wiring Diagram for Krohne Optiflux Sensors and Rosemount 8712 KROHNE ROSEMOUNT 8712...
  • Page 186 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Krohne Generic Sensor Connect coil drive and electrode cables as shown in Figure E-30. to Rosemount 8712 Transmitter Figure E-30. Generic Wiring Diagram for Krohne Sensors and Rosemount 8712 KROHNE SENSORS...
  • Page 187: Siemens Sensors

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 SIEMENS SENSORS Connect coil drive and electrode cables as shown in Figure E-31. Siemens Sitrans F M Mag 3100/5100 Sensors to Rosemount 8712 Transmitter Figure E-31. Wiring Diagram for Siemens Sitrans F M Mag...
  • Page 188: Taylor Sensors

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 TAYLOR SENSORS Connect coil drive and electrode cables as shown in Figure E-32. Series 1100 Sensor to Rosemount 8712 Transmitter Figure E-32. Wiring Diagram for Taylor Series 1100 Sensors and TAYLOR SERIES 1100...
  • Page 189: Taylor Sensors

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Taylor Generic Sensor to Connect coil drive and electrode cables as shown in Figure E-33. Rosemount 8712 Transmitter Figure E-33. Generic Wiring Diagram for Taylor Sensors and ROSEMOUNT 8712 Rosemount 8712...
  • Page 190: Toshiba Sensors

    LF434 Sensor to Rosemount 8712 Transmitter Figure E-34. Generic Wiring Diagram for Toshiba LF430 and ROSEMOUNT 8712 TOSHIBA LF430 LF434 Sensor and Rosemount TRANSMITTER AND LF434 8712 SENSORS Table E-32. Toshiba LF430 and LF434 Sensor Wiring Connections Rosemount 8712 Toshiba LF430 and LF434 Sensors This is a pulsed DC magnetic flowmeter.
  • Page 191: Yamatake Honeywell Sensors

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 YAMATAKE Connect coil drive and electrode cables as shown in Figure E-35. HONEYWELL SENSORS Yamatake Honeywell Sensor to Rosemount 8712 Transmitter Figure E-35. Generic Wiring Diagram for Yamatake ROSEMOUNT 8712 YAMATAKE...
  • Page 192: Yokogawa Sensors

    Connect coil drive and electrode cables as shown in Figure E-36. Yokogawa Sensor to Rosemount 8712 Transmitter Figure E-36. Generic Wiring Diagram for Yokogawa Sensors ROSEMOUNT 8712 YOKOGAWA and Rosemount 8712 TRANSMITTER SENSORS Electrodes Chassis Ground Ex 2 Coils Ex 1 Fuse Table E-34.
  • Page 193: Generic Manufacturer Sensors

    Connect the electrode terminals to Rosemount 8712 terminals 18 and 19. The electrode shield should be connected to terminal 17. Connect the coil terminals to Rosemount 8712 terminals 1, 2, and If the Rosemount 8712 Transmitter indicates a reverse flow condition, switch the electrode wires connected to terminals 8 and 19.
  • Page 194 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 E-40...
  • Page 195: Handheld Communicator

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Appendix F HART Field Communicator Operation HandHeld Communicator ......page F-1 Connections and Hardware .
  • Page 196: Connections And Hardware

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 CONNECTIONS AND The HART Field Communicator exchanges information with the transmitter from the control room, the instrument site, or any wiring termination point in HARDWARE the loop. Be sure to install the instruments in the loop in accordance with intrinsically safe or non-incendive field wiring practices.
  • Page 197: Basic Features

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure F-3. Connecting the HART Field Communicator with NOTE the Optional Load Resistor Loop must be broken to insert the optional 250 ohm resistor. Optional 250 ohm load resistor BASIC FEATURES The basic features of the Handheld Communicator include Action Keys, Function Keys, and Alphanumeric and Shift Keys.
  • Page 198: Alphanumeric And Shift Keys

    00809-0100-4664, Rev BA Rosemount 8712 January 2010 Select “NO” to go to the Main Menu. If a HART-compatible device is found, the communicator displays the Online Menu with device ID (8712) and tag (TRANSMITTER). Page Bksp Delete Page Directional Keys Use these keys to move the cursor up, down, left, or right.
  • Page 199: Fast Key Feature

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 To enter an alphabetic character, first press the Shift key that corresponds to the position of the letter you want on the alphanumeric key. Then press the alphanumeric key. For example, to enter the letter R, first press the right Shift key, then the “6”...
  • Page 200: Online Menu

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 NOTE Online communication with the flowmeter automatically loads the current flowmeter data to the Handheld Communicator. Changes in on-line data are made active by pressing SEND (F2). The transfer function is used only for off-line data retrieval and sending.
  • Page 201: Diagnostic Messages

    Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Diagnostic Messages The following is a list of messages used by the Handheld Communicator (HC) and their corresponding descriptions. Variable parameters within the text of a message are indicated with <variable parameter>.
  • Page 202 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table F-1. Handheld Communicator Diagnostic Messages Message Description No Valid Items The selected menu or edit display contains no valid items. OFF KEY DISABLED Appears when the user attempts to turn the HC off before sending modified data or before completing a method On-line device disconnected with unsent data –...
  • Page 203 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Table F-2. Handheld Fast Keys (HART Handheld Function HART Fast Keys Communicator) and LOI Keys Process Variables 1, 1 Primary Variable (PV) 1, 1, 1 PV Percent of Range 1, 1,2...
  • Page 204 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Function HART Fast Keys Recall Last Saved Values 1,2,3,1,3,3 Set Pass/Fail Criteria 1,2,3,1,4 No Flow Limit 1,2,3,1,4,1 Flowing Limit 1,2,3,1,4,2 Empty Pipe Limit 1,2,3,1,4,3 Measurements 1,2,3,1,5 4-20 mA Verify 1,2,3,2 4-20 mA Verification...
  • Page 205 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Function HART Fast Keys Detailed Setup Additional Parameters 1,4,1 Coil Drive Frequency 1,4,1,1 Density Value 1,4,1,2 PV Upper Sensor Limit (USL) 1,4,1,3 PV Lower Sensor Limit (LSL) 1,4,1,4 PV Minimum Span...
  • Page 206 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Function HART Fast Keys Time Limit 1,4,4,2,4 Coil Drive Frequency 1,4,4,3 Low Flow Cutoff 1,4,4,4 PV Damping 1,4,4,5 Universal Trim 1,4,5 Device Info 1,4,6 Manufacturer 1,4,6,1 1,4,6,2 Descriptor 1,4,6,3 Message 1,4,6,4...
  • Page 207 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 Figure F-6. Field Communicator Menu Tree for the Rosemount 8712 F-13...
  • Page 208 Reference Manual 00809-0100-4664, Rev BA Rosemount 8712 January 2010 F-14...
  • Page 209 Reference Manual 00809-0100-4664, Rev AA Rosemount 8712 January 2010 Index ....A-6 Accuracy Electrical Handheld Communicator ..2-7 .
  • Page 210 Reference Manual 00809-0100-4664, Rev AA Rosemount 8712 January 2010 ... . 2-10 Local Operator Interface (LOI) Pulse Output ..3-6 ..A-4...
  • Page 212 The Emerson logo is a trade mark and service mark of Emerson Electric Co. Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc. PlantWeb is a registered trademark of one of the Emerson Process Management group of companies.

Table of Contents

Save PDF