Doc. No. 0127420 rev. B Table of contents Features ..........................3 Part list..........................4-6 Final device .........................7 How to separate the HI ......................8 Microphone assembly......................9 Hybrid connections ......................10 Coating instructions ......................11 Colour markers replacement .....................12 Test equipment ......................13-14 Charger spares ........................15 Page 2 of 15...
Description Doc. No. 0127420 rev. B Functional Description The ReSound Pulse CRT (Canal Receiver Technology) is a receiver-in-the-ear model in the existing ReSound Pulse family. All features are identical to the sound tube based ReSound models, except from that the receiver is removed from inside the device housing behind the ear and instead placed at the edge of the tube inside the ear canal.
Final Device Doc. No. 0127420 rev. B XREC to be plugged into the stud of the HI Filter removal tool Cerustop filter kit p/n 098-0282 Wax guard Page 7 of 15...
How to separate the HI Doc. No. 0127420 rev. B Pin removal tool p/n 15492900 Gently press out the two pins by using Remove the filter for back microphone the pin removal tool and pull them out by using e.g. tweezers. Take out the battery door and pull out the frame.
Microphone Assembly Doc. No. 0127420 rev. B Wire, amber 12 mm p/n 115129712 Wire, green 12 mm Front mic suspension Back mic suspension p/n 15129812 Wire, red 12 mm p/n 15129612 Make sure that the suspensions fit well to the sound studs/ports on the microphones.
Hybrid connections Doc. No. 0127420 rev. B Battery springs Contact module MIC IN MIC- Coyote 3.0 - No marking Coyote 3.1 - Black dot marking The layout of C3.0 and C3.1 is the same. For visual purpose the C3.1 is Contact marked with black dots.
Coating Doc. No. 0127420 rev. B Cover the pins of the contact module at the sol- der points and at the in-mould. Press it under the contact pins to ensure good protection. Add molykote in the area indicated by the red dotted lines. Molykote Fill the holes at the battery solder points on both sides with Molykote FC1292 p/n 1908-209.
Colour Markers Doc. No. 0127420 rev. B Turn the pin How to place the colour marker: around Place the colour marker as long into the hole as possible. Press on the back of the marker with a finger and at the same time turn the pin around until it breaks of.
Test equipment Doc. No. 0127420 rev. B Blue dot Battery pill must be used: Programming cable CS63 blue, Size 312 p/n 15854900 Left p/n 15206701. Red dot Kit with programming cable, flex strip and Programming cable CS63 red, clothes-clip: Right p/n 15206700. - Red p/n 15254800 - Blue p/n 15254801 - Flex strip 3 pin (10 pcs) p/n 50931703...
Page 14
Test equipment Doc. No. 0127420 rev. B Use XRec test tube p/n 16185100 when testing Snap-on adaptor p/n 15238500. To be used with Otometrics 2cc coupler. Place the Xrec tube into the test fixture. Mount the snap on adaptor on the coupler and place the XRec into it. Connect the hearing instrument to the other end of the Xrec tube The test fixture must be adjusted before use.