Coating Instructions - GN Resound Pulse CRT PS60-R Technical Manual

Table of Contents

Advertisement

Coating
Cover the pins of the contact module at the sol-
der points and at the in-mould. Press it under the
contact pins to ensure good protection.
Fill the holes at the battery solder points on both sides with Molykote FC1292 p/n
1908-209. Wipe the surplus Molykote off to prevent soiling of the housing at final
assembly.
Add molykote in the area indicated by the red dotted lines.
Molykote
Doc. No. 0127420 rev. B
Page 11 of 15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pulse crt lt pss60-r