5.5. Bluetooth

Bluetooth 4.0 BLE
Range 32' (10m) typical and up to 100' (30m) in line of sight
Output power: +0 to -23dBm
Nominal sensitivity: -93dBm
Maximum transfer rate: 10 kbits/s
Average consumption: 3.3µA to 3.3V
5.6. Power Supply
The instrument is powered by three 1.5V LR6 or AA alkaline batteries. You can replace the batteries with
rechargeable NiMH batteries of the same size. However, even when the rechargeable batteries are fully
charged, they will not reach the voltage of the alkaline batteries, and the Battery indicator will appear as
.
Voltage for correct operation is 3 to 4.5V for alkaline batteries and 3.6V for rechargeable batteries. Below 3V,
the instrument stops taking measurements and displays the message BAt.
Battery life (with the Bluetooth connection deactivated) is:
standby mode: 500 hours
recording mode: 3 years at the rate of one measurement every 15 minutes
The instrument can also be powered via a micro USB cable, connected to either a computer or wall outlet
adapter.
5.7. Environmental Conditions
For use indoors and outdoors.
Operating range: +14 to +140°F (-10 to 60°C) and 10 to 90%RH without condensation
Storage range: -4 to +158°F (-20 to +70°C) and 10 to 95%RH without condensation, without batteries
Altitude: <6562' (2000m), and 32,808' (10,000m) in storage
Pollution degree: 2
5.8. Mechanical Specifications
Dimensions (L x W x H): 5.91 x 2.83 x 1.26" (150 x 72 x 32mm)
Mass: 9.17oz (260g) approx.
Inrush protection: IP 50, with the USB connector closed, per IEC 60 529
Drop impact test: 3.28' (1m) per IEC 61010-1
5.9. Compliance with International Standards
The instrument is compliant with standard IEC 61010-1.
5.10. Electromagnetic Compatibility (CEM)
The instrument is compliant with standard IEC 61326-1.
The instruments are not influenced by electromagnetic radiation. However, the sensors for the Models 1821 and
1822 can be affected, due to their wire shape. This can cause them to function as antennas capable of
receiving electromagnetic radiation and affecting measurements.