3.3. Recording Measurements

You can start and stop a recording session on the instrument. Recorded data is stored in the instrument's
memory, and can be downloaded and viewed on a computer running the DataView Data Logger Control Panel.
You can record data by pressing the
A short press (MEM) records the current measurement(s) and date.
A long press (REC) starts the recording session. While the recording is in progress, the symbol REC
appears at the top of the display. A second long press of
while the instrument is recording, a short press of
To schedule recording sessions, and download and view recorded data, consult the DataView Data Logger
Control Panel Help (§4).
3.4. Alarms
You can program alarm thresholds on each measurement channel via the DataView Data Logger Control
Panel. In standalone mode, if an alarm threshold is programmed, the
is crossed, the
symbol blinks, and one of the following blinking symbols appears to the right of the
measurement:
 indicates measurement is above the high threshold.
 indicates measurement is below the low threshold.
 indicates the measurement is between the two thresholds.
3.5. Errors
The instrument detects errors and displays them in the form Er.XX:
Er.01
Hardware malfunction detected. The instrument must be sent in for repair.
Er.02
Internal memory error. Connect the instrument to a computer via the USB cable and format its memory
using Windows.
Er.03
Hardware malfunction detected. The instrument must be sent in for repair.
Er.10
The instrument has not been correctly adjusted. The instrument must be sent to customer service.
Er.11
The firmware is incompatible with the instrument. Install the correct firmware (see §6.4).
Er.12
The firmware version is incompatible with the instrument. Reload the previous firmware version.
Er.13
Recording scheduling error. Ensure that the instrument's time and the time of the DataView Data
Logger Control Panel are the same (see §2.3).
14
button:
has no effect.
Thermometer Data Logger Models 1821, 1822, 1823
stops the recording session. Note that
symbol is displayed. When a threshold

Advertisement