7 PROGRAMMING

Operating Page
Section 6.2, Page 14
xx. x
% Oxygen
xx. x
Alarm Setpoint
Operating Parameters
Secure Parameters

7.1 Access to Secure Parameters

00000
SECURITY CODE
-----

LANGUAGE PAGE

7.2 Language Page
-----
English
Deutsch
Francais
Espanõl
-----
SET UP OUTPUTS
Access to Secure Parameters
Section 7.1, Page 15
00000
SECURITY CODE
Language Page
Section 7.2, Page 15
-----
English
Deutsch
Francais
Espanõl
Fig. 7.1 Overall Programming Chart
Security Code
Enter the required code number, between 00000 and 19999, to gain access to the secure
parameters. If an incorrect value is entered, subsequent programming pages cannot be
accessed.
Advance to
Language
Language Page
Select the language to be displayed on all subsequent pages.
Advance to
Set Up Outputs
Set Up Outputs Page
Section 7.3, Page 16
-----
SET UP OUTPUTS
-----
Alarm Action
EB
EA
5 . 0
Alarm Setpoint
-----
RTX Type
4-20
0-20
0-10
-----
LOG Output
Yes
No
25 . 0 0
Retrans. Span
0 . 2 5
Retrans. Zero
0 . 0
Test Retrans (%)
00000
Alter Sec. Code
page.
page.
-----
ELECTRICAL CAL
-----
Calibrate
YES
NO
xxxxx
mA Zero
(4mA)
xxxxx
mA Span
(20mA)
xxxxx
Res. Zero
(2K0)
xxxxx
Res.Span
(10K0)
-----
Adjust RTX Zero
-----
Adjust RTX Span
15