To Check The Oil Mist Filter - Agilent Technologies 6500 Series Maintenance Manual

Q-tof lc/ms system
Table of Contents

Advertisement

To check the oil mist filter

Check the oil mist filter weekly.
• Check the oil mist filter.
• Check the oil mist filter for damage.
• Check whether oil has collected in the bottom of the oil mist filter.
When you close the ballast valve, you increase the efficiency of the pump. However, you
N O T E
lose oil to the mist filter if you don't recycle. Check the status of your oil mist filter at least
once per week to ensure that it does not fill with oil. If you lose too much oil in the rough
pump, the vacuum will not be maintained, and the Q-TOF LC/MS will vent.
Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
Check for any damage and if pump fluid has been collected in it.
If the oil mist filter is damaged, replace it.
If oil is found in the oil mist filter, open the gas ballast valve counterclock-
wise just enough to return the condensed oil back to the pump. Close the
gas ballast valve clockwise.
Maintenance
To check the oil mist filter
2
141

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents