Defibrillation Pads - Cardiac Science Powerheart G5 User Manual

Automated external defibrillator (aed)
Hide thumbs Also See for Powerheart G5:
Table of Contents

Advertisement

Technical Data
Table 4: Applicable standards (continued)
Type
Detail
Shipping and
ISTA Procedure 2A
transportation
Sensitivity and
• Shockable Rhythm—VF: Meets IEC 60601-2-4
specificity of
requirement and AHA recommendation of
Rhythm Detection
Sensitivity of >90%
• Shockable Rhythm—VT: Meets IEC 60601-2-4
requirement and AHA recommendation of
Sensitivity of >75%
• Non-shockable Rhythm—NSR: Meets IEC
60601-2-4 requirement (>95%) and AHA rec-
ommendation (>99%) of Specificity
• Non-shockable—Asystole: Meets IEC 60601-
2-4 requirement and AHA recommendation of
Specificity of >95%
• Non-shockable—all other rhythms: Meets IEC
60601-2-4 requirement and AHA recommenda-
tion of Specificity – all other rhythms of >95%

Defibrillation pads

Table 5: Preinstalled defibrillation pads (model XELAED001)
Parameter
Detail
Type
Pre-gelled, self-adhesive, disposable, non-polar-
ized (identical pads, which can be placed in either
position) defibrillation pads
Applicable age
Older than 8 years or heavier than 25 kg (55 lb)
and weight of
patient
Shelf life
24 months
Disposal
Check local regulations for disposal information
Table 6: Preinstalled defibrillation pads with CPR feedback
device (model XELAED002)
Parameter
Detail
Type
Pre-gelled, self-adhesive,
disposable, non-polarized (identical pads, which
can be placed in either position) defibrillation
pads with CPR feedback device
Applicable age
Older than 8 years or heavier than 25 kg (55 lb)
and weight of
patient
Shelf life
24 months
Disposal
Check local regulations for disposal information
Powerheart® G5 Automated External Defibrillator
Table 7: Pediatric defibrillation pads (model XELAED003)
Parameter
Detail
Type
Pre-gelled, self-adhesive, disposable, non-polar-
ized (identical pads, which can be placed in either
position) defibrillation pads
Applicable age
Eight years or younger or 25 kg (55 lb) or lighter
and weight of
patient
Shelf life
24 months
Disposal
Check local regulations for disposal information.
Intellisense
battery
®
(model XBTAED001)
Table 8: Intellisense battery
Parameter
Detail
Type
Intellisense lithium battery, non-rechargeable
Output voltage
12 VDC (nominal)
Lithium content
9.2 g (approximate)
Disposal
Check local regulations for disposal information
Estimated shelf
• 5 years from date of manufacture
life*
• Temperature ranges:
Short term (3 days at either temperature
extreme): -30°C to 65°C
Long term (5 years at either temperature
extreme): 20°C to 30°C
Estimated
• Shocks: 420 (typical), 250 (minimum), or 16
operating life**
hours of operating time at 20-30°C, or
(new and fully
Standby: 4 years
charged battery)
* Shelf life is the length of time a battery can be stored prior to installation into an
AED without significantly affecting its operating life.
** The battery operating life depends on the type of battery, device settings, actual
usage, and environmental factors. The number of shocks is estimated at a 300 VE
energy level with a "three shock stack" followed by 60 seconds of CPR using Basic
prompt settings between each set of shocks.
B-3
70-01704-01 B