ECG Analysis Algorithm and Rescue Waveform
Waveform and energy levels for preinstalled
defibrillation pads (Model number XELAED001)
Table C-1: Ultra-low variable energy waveform
(all values are typical)
Phase 1
Patient's
Voltage
Duration
Impedance
(Volts)
(ms)
(Ohms)
(V1)
(T1)
25
1412
3.25
50
1426
4.50
75
1431
5.75
100
1433
7.00
125
1435
8.25
150
1436
9.50
175
1437
10.75
Table C-2: Low variable energy waveform
(all values are typical)
Phase 1
Patient's
Voltage
Duration
Impedance
(Volts)
(ms)
(Ohms)
(V1)
(T1)
25
1631
3.25
50
1647
4.50
75
1653
5.75
100
1655
7.00
125
1657
8.25
150
1658
9.50
175
1659
10.75
Table C-3: High variable energy waveform
(all values are typical)
Phase 1
Patient's
Voltage
Duration
Impedance
(Volts)
(ms)
(Ohms)
(V1)
(T1)
25
1895
3.25
50
1914
4.50
75
1920
5.75
100
1923
7.00
125
1925
8.25
150
1926
9.50
175
1927
10.75
C-2
70-01704-01 B
Phase 2
Voltage
Duration
Energy
(Volts)
(ms)
(Joules)
(V2)
(T2)
743
3.2
146-197
907
3.2
128-172
968
3.2
116-156
1000
3.2
108-144
1019
3.2
102-136
1031
3.2
97-130
1038
3.2
94-126
Phase 2
Voltage
Duration
Energy
(Volts)
(ms)
(Joules)
(V2)
(T2)
858
3.2
195-263
1047
3.2
170-230
1118
3.2
154-208
1155
3.2
143-193
1176
3.2
135-182
1190
3.2
129-174
1199
3.2
125-168
Phase 2
Voltage
Duration
Energy
(Volts)
(ms)
(Joules)
(V2)
(T2)
997
3.2
263-355
1216
3.2
230-310
1299
3.2
208-280
1342
3.2
193-260
1367
3.2
183-246
1383
3.2
174-235
1393
3.2
168-226
Waveform and energy levels for pediatric
defibrillation pads
Table C-4: Ultra-low variable energy waveform
(all values are typical)
Phase 1
Patient's
Voltage
Duration
Impedance
(Volts)
(ms)
(Ohms)
(V1)
(T1)
25
682
3.25
50
689
4.50
75
691
5.75
100
692
7.00
125
693
8.25
150
694
9.50
175
694
10.75
Table C-5: Low variable energy waveform
(all values are typical)
Phase 1
Patient's
Voltage
Duration
Impedance
(Volts)
(ms)
(Ohms)
(V1)
(T1)
25
791
3.25
50
798
4.50
75
801
5.75
100
802
7.00
125
803
8.25
150
804
9.50
175
804
10.75
Table C-6: High variable energy waveform
(all values are typical)
Phase 1
Patient's
Voltage
Duration
Impedance
(Volts)
(ms)
(Ohms)
(V1)
(T1)
25
915
3.25
50
924
4.50
75
927
5.75
100
929
7.00
125
930
8.25
150
931
9.50
175
931
10.75
Powerheart® G5 Automated External Defibrillator
Phase 2
Voltage
Duration
Energy
(Volts)
(ms)
(Joules)
(V2)
(T2)
359
3.2
35-46
438
3.2
30-40
468
3.2
27-36
483
3.2
25-33
493
3.2
24-31
498
3.2
23-30
802
3.2
22-29
Phase 2
Voltage
Duration
Energy
(Volts)
(ms)
(Joules)
(V2)
(T2)
416
3.2
46-61
508
3.2
40-54
542
3.2
37-48
560
3.2
34-45
570
3.2
32-42
577
3.2
31-40
581
3.2
30-39
Phase 2
Voltage
Duration
Energy
(Volts)
(ms)
(Joules)
(V2)
(T2)
481
3.2
62-82
588
3.2
54-72
628
3.2
49-65
648
3.2
46-60
660
3.2
43-57
668
3.2
41-54
673
3.2
40-52